QIAcube Connect MDx User Manual 12/2020
77
A
1.5 ml microcentrifuge
tube lid is in correct
position
B
Correctly loaded rotor adapter.
A
Rotor adapter is correctly loaded, and the 1.5 ml microcentrifuge tube lid is in the
correct position;
B
Correctly loaded rotor adapter seen from the side.
C
1.5 ml microcentrifuge
tube lid not pushed
down into slot
D
Incorrectly loaded rotor adapter. C
Rotor adapter is incorrectly loaded with a 1.5 ml microcentrifuge tube. The lid of the
tube is not pushed all the way down to the bottom of the slot of the rotor adapter and could break off during centrifugation
(compare with part A of the above figure);
D
Incorrectly loaded rotor adapter seen from the side (compare with part
B
of
the above figure).
E
1.5 ml microcentrifuge
tube lid is in the wrong
slot of the rotor adapter
Rotor adapter is incorrectly loaded with a 1.5 ml microcentrifuge tube.
The lid of the tube is positioned in the wrong slot of
the rotor adapter. During column transfer, the lid of the spin column could crash onto the lid of the 1.5 ml microcentrifuge
tube, causing the protocol run to crash.