QIAcube Connect MDx User Manual 12/2020
28
WARNING
Electrical hazard
[W8]
Any interruption of the protective conductor (earth/ground lead) inside or
outside the instrument or disconnection of the protective conductor terminal is
likely to make the instrument dangerous.
Intentional interruption is prohibited.
Lethal voltages inside the instrument
When the instrument is connected to line power, terminals may be live and
opening covers or removing parts is likely to expose live parts.
WARNING
Damage to electronics
[W9]
Before powering ON the instrument, make sure that the correct supply voltage
is used.
Use of incorrect supply voltage may damage the electronics.
To check the recommended supply voltage, refer to the specifications
indicated in the type plate of the instrument.
WARNING
Risk of electric shock
[W10]
Do not open any panels on the QIAcube Connect MDx.
Risk of personal injury and material damage
Only perform maintenance that is specifically described in this user manual.
Cooling air outlet
Cooling air outlets are located at the rear left side of the QIAcube Connect MDx and allow cooling
of the internal components of the QIAcube Connect MDx.
CAUTION
Risk of overheating
[C9]
To ensure proper ventilation, maintain a minimum clearance of 10 cm at the
sides and rear of the QIAcube Connect MDx.
Slits and openings that ensure the ventilation of the QIAcube Connect MDx
must not be covered.