135
O-Ring tool with prepared new O-Ring
The O-Ring replacement is semi-automatically and includes movement of robotic arm.
Note
: Perform the daily maintenance and cleaning of the robotic arm procedure before you replace
the O-Ring.
1.
To start the O-Ring replacement, press the
Tools
icon on the menu bar. Press the
Maintenance
tab and select the
O-Ring
subtab. The screen shows the last O-Ring replacement date.
O-Ring maintenance screen
2.
Close the hood and press
Start
to begin the O-Ring replacement procedure. Follow the
instructions on the screen. Details are provided in the next steps below.