QIAcube Connect MDx User Manual 12/2020
17
CAUTION
Damage to the instrument
[C8]
After a power failure, do not move the z-module (robotic arm) manually in
front of the instrument. Damage may occur if the QIAcube Connect MDx hood
is closed and collides with the z-module.
WARNING
Risk of personal injury and material damage
[W20]
Raise the centrifuge lid carefully. The lid is heavy and may cause injury if it
falls.
CAUTION
Risk of overheating
[C9]
To ensure proper ventilation, maintain a minimum clearance of 10 cm at the
sides and rear of the QIAcube Connect MDx.
Slits and openings that ensure the ventilation of the QIAcube Connect MDx
must not be covered.
2.8
Heat hazard
The QIAcube Connect MDx worktable contains a heated shaker.
WARNING
Hot surface
[W21]
The shaker can reach temperatures of up to 70°C (158°F). Avoid touching it
when it is hot. Carefully remove the samples after a run.
2.9
Maintenance safety
WARNING/
CAUTION
Risk of personal injury and material damage
[W22]
Only perform maintenance that is specifically described in this user manual.