QIAcube Connect MDx User Manual 12/2020
45
4.4
Configuration of the QIAcube Connect MDx
When using the QIAcube Connect MDx for the first time, it is recommended to define the required
settings. Other settings can be made later when needed.
For details on using the touchscreen and software, refer to section 5.1, QIAcube Connect MDx
software.
To configure the QIAcube Connect MDx, follow the steps below.
1.
Close the instrument hood.
2.
Press the power switch to the inner position to power ON the instrument. The startup screen
appears and the beeper sounds (if enabled in the sound settings). The instrument automatically
performs the initialization tests. If the centrifuge lid is closed, it will open.
3.
Initially, only one user account is available: the pre-installed default user. Press
OK
on the
touchscreen to confirm the message.
4.
Initially, if no maintenance has been recorded yet, the maintenance status is initialized using a
default file. Press
OK
to confirm the message.
5.
The
Login
screen appears after the initialization.
Login screen
6.
Initially, only one default user is available. In this case, enter
Admin
in both the
User ID
and
Password
fields using the on-screen keyboard. Touch in the entry field to open the on-screen
keyboard. For details, refer to section 5.1, QIAcube Connect MDx software, on how to enter
text or numbers. In addition, please select which software mode (IVD or Research) to launch.