QIAcube Connect MDx User Manual 12/2020
32
Reagent bottle rack with white labeling strips at the sides
WARNING
Risk of fire or explosion
[W6]
When using ethanol or ethanol-based liquids on the QIAcube Connect MDx,
handle such liquids carefully and in accordance with the required safety
regulations. If liquid has been spilled, wipe it off and leave the QIAcube
Connect MDx hood open to allow flammable vapors to disperse.
WARNING
Risk of explosion
[W7]
The QIAcube Connect MDx is intended for use with reagents and substances
supplied with QIAGEN kits or other than outlined in respective Information
for use. Use of other reagents and substances may lead to fire or explosion.
WARNING
Risk of personal injury and material damage
[W5]
To prevent plastic crash, load the tubes properly. After a plastic crash, sharp
plastic particles could be inside the centrifuge. Be careful when handling
items inside the centrifuge.
WARNING
Hot surface
[W21]
The shaker can reach temperatures of up to 70°C (158°F). Avoid touching it
when it is hot. Carefully remove the samples after a run.