QIAcube Connect MDx User Manual 12/2020
12
WARNING
Damage to electronics
[W9]
Before powering ON the instrument, make sure that the correct supply voltage
is used.
Use of incorrect supply voltage may damage the electronics.
To check the recommended supply voltage, refer to the specifications
indicated in the type plate of the instrument.
WARNING
Risk of electric shock
[W10]
Do not open any panels on the QIAcube Connect MDx.
Risk of personal injury and material damage
Only perform maintenance that is specifically described in this user manual.
To ensure satisfactory and safe operation of the QIAcube Connect MDx, follow these guidelines:
The line power cord must be connected to a line power outlet that has a protective conductor
(earth/ground).
Do not adjust or replace internal parts of the instrument.
Do not operate the instrument with any covers or parts removed.
If liquid has spilled inside the instrument, power OFF the instrument, disconnect it from the
power outlet, and contact QIAGEN Technical Services.
If the instrument becomes electrically unsafe, prevent other personnel from operating it, and contact
QIAGEN Technical Services.
The instrument may be electrically unsafe when:
It or the line power cord appears to be damaged.
It has been stored under unfavorable conditions for a prolonged period.
It has been subjected to severe transport stresses.
Liquids come in contact directly with electrical components of the QIAcube Connect MDx.
2.3
Environment
Parameters such as temperature range and humidity range are described in section 9,
Appendix A – Technical data.