QIAcube Connect MDx User Manual 12/2020
72
Example of message box displayed by hitting the information icon (
).
2.
Make sure that the buffer bottles contain the minimum volumes described in the
Volume
column. Each bottle can hold a maximum volume of 30 ml, but it is recommended not to use
more than the minimal volume.
Later, when the run is started, the instrument will determine the filling volume. Ensure to label
the buffer bottles properly and according to safety requirements. Buffer bottles can be stored
according to the storage conditions described in the kit handbooks. However prolonged times
of buffer bottles standing open on the instrument must be avoided. For subsequent runs, fresh
buffer should be refilled. We recommend to reuse the buffer bottles only until one kit is used
up. As soon as a new QIAGEN kit is opened, new buffer bottles should be used.
3.
Place each opened buffer bottle in the correct position of the reagent bottle rack as shown on
the screen. The buffer bottle rack positions are numbered for easy identification. The rack will
only fit on the worktable if its orientation is correctly set.
4.
Once all the buffer bottles are placed on the reagent bottle rack, place the rack on the
worktable. Ensure that the rack is oriented correctly with number 1 on the top.
Important
: Ensure to place the reagent bottle rack correctly into the designated worktable slot.
Bottle racks that are in a leaning position could cause errors during liquid detection.
Important
: Ensure that buffer bottles are open. A closed buffer bottle will be detected by the
instrument and prevent the run from being started.
5.
Press
Next
to proceed to loading the tips and enzymes. The software automatically proceeds
to the next screen. Follow the instructions in the corresponding section below.