QIAcube Connect MDx User Manual 12/2020
87
the software version
the sample IDs and their positions
the time, date, and user at the run start
the time and date of the run end
the user that confirmed the end of run
the kit material number, lot number, and expiration date
the error and warning descriptions
the run validity (valid or invalid)
the run status (completed or aborted)
the software mode (IVD or Research)
the run ID
the elution volume
the final eluate position
Important
: It is recommended to perform regular maintenance as described in section 6.3,
Regular maintenance procedure, before starting the next run.
5.6
Stopping a protocol run
In case of emergency, a run can be stopped by pressing the
Stop Run
button on the run status
screen (see section 5.5, Starting a protocol run). To confirm the run stopped, click
Yes
in the
Stop
Run
dialog.
Stop run screen