PULSAR MEASUREMENT
189
Important Notice
When mA Output 1 is to be representative of the
average
or
sum
of two
points of measurement, then
both points
must be set to the
same
units
of measurement. In the case of
flow P*100
and
P*100
are set for
4
(OCM Head)
or
5 (OCM Flow)
, then
P*70
6 (
Volume Units
) &
P*707
(
Time Units
) must be the same. And in the case of
Volume
then
P*100
are set for
6 (Volume)
then
P*605 (Volume Units)
must be the same.
Important Notice
When
both
mA Output 1
and
mA Output 2
are
allocated
to the
same
point
of measurement, for them to output the same reading, both
mA
low value
(
P*834/P*892
) and
mA high value
(
P*835/P*893
) must be
the
same
for each mA Output.