PULSAR MEASUREMENT
163
OCM
P1 and P2
:
When P*100 = 4 (OCM Head) or 5 (OCM Flow)
PMD Setup
P*700 Primary Measuring Device
This parameter is used to select the
type
of
Primary Measuring Device
and
enable additional parameters required to calculate the flow of the Primary
Measuring Device chosen (P*701). Options are as follows:
0 = Off (Default)
1 = Exponent
2 = BS3680 Flume
3 = BS3680 Weir
4 = Not Available
5 = Special
6 = Universal
P*701 Primary Measuring Device
Enter the Primary Measuring Device used.
If P*700 = 1 (Exponent)
Select from the following options:
1 = Suppressed Rectangular Weir
2 = Cipolletti (Trapezoidal) Weir
3 = Venturi Flume
4 = Parshall Flume
5 = Leopold Lagco Flume
6 = V- notch Weir
7 = Other
If P*700 = 2 (BS 3680 Flume)
Select from the following options:
1 = Rectangular
2 = Rectangular with hump
3 = U-throated
4 = U-Throated with hump