88
SECTION 4
When SECS Is
N
ot U
sed
CIDRW System
User’s Manual
SECTION 4
Reading from/Writing to ID Tags
The command and response details are given for a 1:N protocol. Just as in the previous command format, the
details for a 1:1 protocol are the same if the SOH, node number, and FCS are deleted.
READ
Reads any pages of data from the ID Tag. The maximum number of pages that can be read at one time
is 16.
ID Tag Memory Maps
Refer to page 159.
The response code (when normal: 00) and the data in the specified pages are returned in ascending order of
page numbers.
Parameter Description
Parameter
Description
Page designation
Pages are specified by setting the bits corresponding to pages that are to be read to 1 and setting
the other bits to 0, then converting the result to a hexadecimal character string.
SOH Node No.
Response
code
Read data
FCS
CR
Page n
⋅ ⋅ ⋅
Page m (n<m)
Data 1
⋅ ⋅ ⋅
Data 8
Data 1
⋅ ⋅ ⋅
Data 8
01h
0
0
0Dh
Command
Bit
7
-
0
7
-
3
2
1
0
7
6
-
1
0
7
6
-
2
1
0
Page
Sys
-
Sys Sys
-
Sys
17
16
15
14
13
-
8
7
6
5
-
1
Sys Sys
Designation
0*
0*
0*
0*
0*
0/1
0/1
0/1
0/1
0/1
•••
0/1
0/1
0/1
0/1
•••
0/1
0*
0*
Value
00
00 to 07
00 to FF
00 to FC
* Always specify 0. If you specify 1 an error (Response code: 14) will occur.
SOH
Node No.
Command code
Page designation (8 characters)
FCS
CR
01h
0
1
0
0
0Dh
Response
Содержание V640-HS61
Страница 14: ...12 INTRODUCTION Reading this Manual CIDRW System User s Manual INTRODUCTION MEMO...
Страница 76: ...74 SECTION 3 Communications Test CIDRW System User s Manual SECTION 3 Preparing for Communications...
Страница 106: ...104 SECTION 4 When SECS Is Not Used CIDRW System User s Manual SECTION 4 Reading from Writing to ID Tags...
Страница 120: ...118 SECTION 5 When SECS Is Not Used CIDRW System User s Manual SECTION 5 Troubleshooting MEMO...
Страница 171: ...CIDRW System User s Manual SECTION 6 Protective Construction SECTION 6 Appendix 169 MEMO...
Страница 173: ......