55
CIDRW System
User’s Manual
SECTION 3
Set
ting t
h
e Communica
tions
Cond
it
ions for the
CIDR
W
Controll
er
SECTION 3
Preparing for Communications
1.
Specify the parameters to be changed.
When the first parameter is specified, the ALARMS indicator flashes.
2.
Confirm the parameter change.
The input parameter is checked and written.
When writing is completed, a message indicating the result is displayed.
The ALARMS indicator lights.
If writing is completed with an error, the parameters are not updated.
The figure in square brackets [ ] indicates the line number where the
error was first detected. If a parity error is detected in the received char-
acters, this figure is [0].
Check the sent data based on this information.
A text file is created based on the data that is keyed in, as shown below, and this data can be conveniently transmitted
using the terminal's text file send function.
Check for Correct Setting
The currently set data can be output so that you can check if it is correct.
1.
Send the parameter output command "::GET_PARAM"
from the host device.
S_BAUD=19200
S_DEVID=1
S_BNO=0
_
::END
_
SETUP_COMPLETE
_
SETUP_FAILED [2]_
When writing is completed without error
When writing is completed with an error
#Parameter Setting File for SystemA
#Protocol
S_BAUD=19200
S_DEVID=1
#SECS
S_BNO=0
::END
Example: PRM.TXT
::GET_PARAM
Содержание V640-HS61
Страница 14: ...12 INTRODUCTION Reading this Manual CIDRW System User s Manual INTRODUCTION MEMO...
Страница 76: ...74 SECTION 3 Communications Test CIDRW System User s Manual SECTION 3 Preparing for Communications...
Страница 106: ...104 SECTION 4 When SECS Is Not Used CIDRW System User s Manual SECTION 4 Reading from Writing to ID Tags...
Страница 120: ...118 SECTION 5 When SECS Is Not Used CIDRW System User s Manual SECTION 5 Troubleshooting MEMO...
Страница 171: ...CIDRW System User s Manual SECTION 6 Protective Construction SECTION 6 Appendix 169 MEMO...
Страница 173: ......