When SECS Is Not Used
CIDRW System
User’s Manual
SECTION 5
Troubleshooting
• Amplifier Unit Error
Check the status of the indicators after transmission of the test command.
After taking appropriate corrective action, restart the Amplifier Unit, send the test command again and
check again.
• If There Is No Response to the Command:
Check the status of the indicators after transmission of the test command.
After taking appropriate corrective action, restart the Amplifier Unit, send the test command again and
check again.
Using RS Signal Control at the Host Device
In a 1:N connection using Link Units, the RS signals generated from the host device by normal control must be input as
CS signals. Turn the RS signals OFF within 15 ms after the completion of data transmission. Correct communications
will not be possible without this control. When using a USB-serial adapter, direct control of the RS signal may not be
possible. Test operation in advance to make sure direct control of the RS signal is correct.
• If There Is a Response to the Command:
Check the status of the indicators after transmission of the test command.
After taking appropriate corrective action, restart the Amplifier Unit, send the test command again and
check again.
RUN
COMM
NORM
ERROR
Main check points
—
(If RUN is OFF, the status of the other indica-
tors can be ignored.)
• Influence of background noise (change installation position)
• Amplifier Unit power supply
If the error cannot be resolved after checking, the Amplifier Unit
may be damaged.
RUN
COMM
NORM
ERROR
Main check points
• Amplifier Unit baud rate settings
• Node numbers of the Amplifier Units (do not match the node
number in the test command)
• Connection and wiring of the cable between the host device
and Amplifier Unit
• OFF timing of the RS signals between the host device and
Amplifier Unit
• Routing of each cable (influence of background noise)
If the error cannot be resolved after checking, the Amplifier Unit
may be damaged.
• Amplifier Unit baud rate settings
• Amplifier Unit node number setting (More than one Amplifier
Unit may be set to the same node number.)
• Connection and wiring of the cable between the host device
and Amplifier Unit
• Routing of the cables (influence of background noise)
• OFF timing of the RS signals between the host device and
Amplifier Unit
• FCS (frame check sequence) calculation method
RUN
COMM
NORM
ERROR
Main check points
• Node numbers of the Amplifier Units (The same number is set
for more than one Unit)
If the error cannot be resolved after checking, the Amplifier
Unit may be damaged.
There is a mistake in the command format (number of charac-
ters, character code, etc.).
(Lights once)
SD at host device
RS at host device
ON only during data transmission from the host device
Within 15 ms
(Lights once)
Содержание V640-HS61
Страница 14: ...12 INTRODUCTION Reading this Manual CIDRW System User s Manual INTRODUCTION MEMO...
Страница 76: ...74 SECTION 3 Communications Test CIDRW System User s Manual SECTION 3 Preparing for Communications...
Страница 106: ...104 SECTION 4 When SECS Is Not Used CIDRW System User s Manual SECTION 4 Reading from Writing to ID Tags...
Страница 120: ...118 SECTION 5 When SECS Is Not Used CIDRW System User s Manual SECTION 5 Troubleshooting MEMO...
Страница 171: ...CIDRW System User s Manual SECTION 6 Protective Construction SECTION 6 Appendix 169 MEMO...
Страница 173: ......