5.8 Measurement Accuracy
5
-
22
ZA57630
5.8.2 Measurement Mode IMPD
−3T
Unless otherwise specified, the conditions are that 0 to +40 °C, open and short correction was
performed after warming up for at least 30 minutes.
・
Basic accuracy
±0.08 %
・
Measurement range
Measurement
range Z
R
Measurable
range
Recommended range
1 MΩ
900 kΩ ≤
1 MΩ to 11 MΩ
100 kΩ
90 kΩ ≤
100 kΩ to 1.1 MΩ
10 kΩ
9 kΩ ≤
10 kΩ to 110 kΩ
1 kΩ
900 Ω ≤
1 kΩ to 11 kΩ
100 Ω
No limitation
9 Ω to 1.1 kΩ
10 Ω
≤ 10 Ω
1 Ω to 10 Ω
Measurable range:
Approximate range in which measurement
and display are possible (supplementary
value)
Recommended range: Operating range in which measurement
accuracy is high.
・
Impedance measurement accuracy
Accuracy of |Z|
±A
Z
[%]
A
Z
=
{(A
+
B×U
+
K
Z
+
K
Y
)×K
V
+
K
B
}×K
T
Accuracy of phase angle θ of impedance
±P
Z
[°]
when 10 kHz < f < 30 kHz and measurement range is 1 kΩ
P
Z
=
0.573×{(1.5×A
+
1.5×B×U
+
K
Z
+
K
Y
)×K
V
+
K
B
}×K
T
when 10 kHz < f < 30 kHz and measurement range is 100 Ω
P
Z
=
0.573×{(2×A
+
2×B×U
+
K
Z
+
K
Y
)×K
V
+
K
B
}×K
T
other than above
P
Z
=
0.573×A
Z
Remark
The measurement accuracy when A
Z
exceeds 10 % is a
supplementary value.
Excluding the highest and lowest measurement ranges that can be
used with that frequency, the measurement accuracy for a
measured value smaller than half the lower limit of each
recommended measurement range or larger than twice the upper
limit is a supplementary value.
Each parameter value in the expression of A
Z
and P
Z
is listed below. The meaning of the
symbol used when calculating each parameter is shown below.
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...