3.5 Setting Screen Reference
3
-
101
ZA 57630
For ideal measurement data, the result will be the same no matter which method you select.
However, the results obtained by the above four algorithms will differ for actual measurement
data containing noise, measurement errors, and other problems. Select the C
0
estimation mode
in reference to the equivalent circuit fit results. Usually, “RESNf” should be used.
■ [PZT]
-
[PZT PRMTR]
-
[FREQ RESOLN]
:
Set frequency resolution for impedance characteristic simulation.
○ Setting procedure
<Numerical settings (requiring confirmation)>, <Numerical settings>
○ Functions of function keys
・
[SWEEP SETTING]
: It will set the same value as the sweep resolution setting value.
・
[EDIT]
: Set the frequency resolution with the numerical input palette.
・
[MODIFY]
: Modify the frequency resolution with the knob or cursor keys.
The impedance characteristic is calculated from the equivalent circuit estimation results and
then displayed in a graph. The impedance characteristic is calculated with a number of
frequency points of the value set here + 1 in the frequency resolution of this time and then
displayed on the graph.
■ [PZT]
-
[PZT CONST]:
Calculate the piezoelectric constants.
■ [PZT]
-
[PZT CONST]
-
[[PZT CONSTANT]]
: Display the piezoelectric constant calculation dialog box.
This displays a dialog box to enter information such as the DUT piezoelectric ceramic shape
and vibration mode to calculate various piezoelectric constants.
■ [REL PERMT]:
Set parameters for relative permittivity measurement.
This sets the shape information of the DUT for displaying the relative permittivity of the
DUT. It can be set when the measurement mode is IMPD−EXT, IMPD−3T and IMPD−2T.
When relative permittivity related (ε
S
, ε
S
', ε
S
'', or D
ε
) measurement is selected in
[Graph]
-
[DISPLAY MODE Y]
or
[Measure]
-
[PRMTR
n
], the relative permittivity will be calculated
using the parameters that you set here.
■ [REL PERMT]
-
[ELECTRODE AREA]:
Set the electrode surface area.
○ Setting procedure
<Numerical settings (requiring confirmation)>, <Numerical settings>
○ Functions of function keys
・
[EDIT]
: Set the electrode surface area (unit: mm
2
) with the numerical input palette.
・
[MODIFY]
: Modify the electrode surface area (unit: mm
2
) with the knob or cursor keys.
■ [REL PERMT]
-
[ELECTRODE DISTANCE]
:
Set the distance between electrodes.
○ Setting procedure
<Numerical settings (requiring confirmation)>, <Numerical settings>
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...