3.5 Setting Screen Reference
3
-
95
ZA 57630
■ [PORT EXT]
-
[LOAD_PEXT CORR]
-
[FUNCTION]
:
Enable or disable load correction function at port extension tip.
Enable or disable the load correction function at port extension tip.
○ Setting procedure
<Numerical settings (requiring confirmation)>, <Numerical settings>
○ Functions of function keys
・
[ON/OFF]
: Enable or disable load correction at port extension tip.
・
[MEASURE]
: Start correction data measurement (when measurement mode is
IMPD−3T and IMPD−2T only).
・
[COPY TO REF8]
: Copy the load correction data at port extension tip to reference
data 8 (REF8).
・
[EDIT]
: Set the load correction data number at port extension tip with the
numerical input palette.
・
[MODIFY]
: Modify the load correction data number at port extension tip with
the knob or cursor keys.
The currently set load correction data number at port extension tip is displayed in the menu.
When load correction at port extension tip is turned off, the data number indication becomes
“OFF”.
When load correction at port extension tip is enabled, this instrument will operate with open
correction and short correction at port extension tip both forcibly enabled internally even if
those functions were set to off (disabled).
■ [PORT EXT]
-
[LOAD_PEXT CORR]
-
[LOAD STD FMT]
:
Set the load standard value format at port extension tip.
This specifies the format for when the load standard values at port extension tip are input.
○ Setting procedure
<List operations (immediate settings)>
・
Cs-D
: Specify with the series capacitance and dissipation factor.
・
Cp-D
: Specify with the parallel capacitance and dissipation factor.
・
Rp-Cp
: Specify with the parallel resistance and parallel capacitance.
・
Rs-Ls
: Specify with the series resistance and series inductance.
・
Rs-X
: Specify with the resistance (series resistance) and reactance.
・
Z-
θ
: Specify with the impedance and phase.
■ [PORT EXT]
-
[LOAD_PEXT CORR]
-
[LOAD STD VALUE]:
:
Set the load standard values at port extension tip.
This sets the load standard values at port extension tip.
When the menu is tapped, the
[[PORT EXTENSION LOAD STANDARD VALUE]]
dialog box
appears. Set the frequency and load standard values at port extension tip. Up to 30 sets of
load standard values (30 frequencies) at port extension tip can be set.
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...