4.5 Setting Sweep Measurement
4
-
20
ZA57630
The unit for the upper and lower limits differs depending on the sweep item (
[Sweep]
−
[SWEEP ITEM]
) setting.
● Sweep range when zero-span sweep
The sweep range cannot be set for a zero-span sweep. The sweep lower limit is fixed to 0 s.
The sweep time is (the time required to perform one measurement (measurement time
setting)) × (the sweep density setting (
[Sweep]
− [SWEEP RESOLUTION])
). When a zero-span
sweep is performed, the sweep upper time limit (sweep end time) may become even longer if
auto range, automatic high density sweep, or ALC is set.
■ Setting the display mode
Select the items to display on the X-axis, Y1-axis, and Y2-axis of the graph in
[Graph]
−
[DISPLAY MODE X]
/ [DISPLAY MODE Y]
. The items that can be selected differ depending on
the measurement mode. If a setting other than one in this table is set, nothing will be
displayed in the graph. LIN indicates the linear axis, and LOG the logarithmic axis.
Furthermore, “
” and “×” indicate whether or not the linear axis or logarithmic axis can be set.
Table 4-1
Display Modes (Measurement Mode: IMPD−EXT / IMPD−3T / IMPD−2T)
X-axis
LIN LOG Y1-axis LIN LOG Y2-axis LIN LOG
Remarks
SWEEP
*1
Z
θ
Z
×
SWEEP
*1
Y
θ
Y
×
SWEEP
*1
R
×
X
×
SWEEP
*1
G
×
B
×
SWEEP
*1
C
P
×
R
P
×
SWEEP
*1
C
P
×
D
×
SWEEP
*1
C
P
×
Q
C
×
SWEEP
*1
C
S
×
R
S
×
SWEEP
*1
C
S
×
D
×
SWEEP
*1
C
S
×
Q
C
×
SWEEP
*1
L
P
×
R
P
×
SWEEP
*1
L
P
×
Q
L
×
SWEEP
*1
L
S
×
R
S
×
SWEEP
*1
L
S
×
Q
L
×
SWEEP
*1
ε
S
×
D
ε
×
Relative permittivity
SWEEP
*1
ε
S
'
×
ε
S
''
×
Relative permittivity
SWEEP
*1
µ
S
×
D
µ
×
Relative permeability
SWEEP
*1
µ
S
'
×
µ
S
''
×
Relative permeability
SWEEP
*1
V1
V2
Voltage monitoring
*2
SWEEP
*1
V
I
Voltage/current monitoring
*3
SWEEP
*4
×
FREQ
×
θ
Z
×
Resonant frequency
tracking measurement
R
×
X
×
—
—
—
Complex impedance plane
R
×
-X
×
—
—
—
Cole-cole plot
G
×
B
×
—
—
—
Complex admittance plane
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...