4.6 Setting Spot Measurement
4
-
25
ZA57630
4.6 Setting Spot Measurement
This section describes the basic settings for spot measurement.
Up to six parameters can be simultaneously measured and displayed in spot measurement.
In addition, unlike sweep measurement, there are no limitations on the parameters that can
be displayed simultaneously.
Set the parameters to display in spot measurement in
[Measure]
− [PRMTR1]
to
[PRMTR6]
.
Tap the menu to display a palette listing the parameters that can be measured. Tap to select the
parameters to measure and display in the palette. The measurement and display items can also
be changed by tapping the parameter buttons in the spot measurement result screen.
If you select
[NA]
, that parameter will be hidden (only PRMTR3 to 6). The types and formulas
of parameters other than
[STATUS]
are the same as for sweep measurement. Refer to Table
4-3 and Table 4-6.
[STATUS]
displays information such as the range and over-level information for measurement.
(Measurement Mode: IMPD−EXT and G−PH)
AUTO, 100mV, 50mV, 0
Over-level information 0 to 15
PORT2 measurement range
PORT1 measurement range
FIX
:
When both PORT1 and PORT2 are fixed range
AUTO
:
When any of PORT1 and PORT2 are auto range
The over-level information is a decimal number from 0 to 13 with the following value added.
When PORT1 over
+8
When PORT2 over
+4
When ALC error
+1
(When measurement mode is IMPD−3T or IMPD−2T)
(When measurement mode is IMPD−3T or IMPD−2T)
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...