3.5 Setting Screen Reference
3
-
45
ZA 57630
・
[A]
: Set with current (Arms, A).
■ [AMPLITUDE]
:
Set the measurement signal amplitude.
This sets the measurement signal amplitude for performing spot measurement.
○ Setting procedure
<Numerical settings (immediate settings)>, <Numerical settings>
○ Functions of function keys
・
[SWEEP MIN -> AMPTD]
*
: Copy the sweep lower amplitude limit to
AMPLITUDE
.
・
[SWEEP MAX -> AMPTD]
*
: Copy the sweep upper amplitude limit to
AMPLITUDE
.
・
[MKR -> AMPTD]
*
: Copy the amplitude of the active marker to
AMPLITUDE
.
・
[CURR -> AMPTD]
: Copy the current output amplitude to
AMPLITUDE
.
・
[EDIT]
: Set the amplitude with the numerical input palette.
・
[MODIFY]
: Modify the amplitude setting with the knob or cursor keys.
*: Displayed when measurement signal amplitude sweep.
■ [AMPLITUDE LIMIT]
:
Set the amplitude limit.
This function is to prevent excessive output due to an incorrect measurement signal
amplitude setting and prevent excessive output when ALC is enabled. An measurement
signal amplitude above that set here will not be output even when ALC is enabled.
○ Setting procedure
<Numerical settings (immediate settings)>, <Numerical settings>
○ Functions of function keys
・
[SWEEP MAX -> LIMIT]
*
: Copy the sweep upper amplitude limit to
AMPLITUDE LIMIT
.
・
[MKR -> LIMIT]
*
: Copy the amplitude of the active marker to
AMPLITUDE LIMIT
.
・
[EDIT]
: Set the amplitude limit with the numerical input palette.
・
[MODIFY]
: Modify the amplitude limit with the knob or cursor keys.
*: Displayed when measurement signal amplitude sweep.
■ [ALC]
:
Set automatic level control.
This is used to automatically adjust the measurement signal output amplitude for supplying a
constant amplitude to the DUT. For details, refer to
“1.4.4 ALC”
in
Instruction Manual (Advanced)
.
■ [ALC]
-
[ALC]
:
Enable or disable.
This sets whether or not to perform the ALC operation.
○ Setting procedure
<List operations (immediate settings)>
・
OFF
: Disable the ALC.
・
CV1
*1
: Control the amplitude so that the PORT1 input amplitude becomes constant.
・
CV2
*1
: Control the amplitude so that the PORT2 input amplitude becomes constant.
・
ON
*2
: Enable the amplitude compression function (constant voltage or constant current).
*1: Displayed only in the IMPD−EXT and G−PH measurement modes.
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...