5.7 Measured Signal Control Section
5
-
18
ZA57630
5.7 Measured Signal Control Section
・
Signal output control
Measurement synchronous drive
SYNC(AC+DC) The measurement signal and DC bias are turned on at the start of
measurement and turned off at the end of measurement.
SYNC(AC)
The measurement signal is turned on at the start of measurement
and turned off at the end of measurement. The DC bias does not
change.
ASYNC
The measurement signal and DC bias are not changed at the start
of measurement and end of measurement.
ON/OFF mode
QUICK
The measurement signal amplitude and DC bias changes
immediately.
SLOW
Output changes gradually over a period of approximately 10
seconds.
0° SYNC
This instrument waits until the measurement signal phase
becomes 0° and then output turns off
Remark
Regardless of the setting, output always turns on at phase 0°.
Frequency change mode
ASYNC
The frequency changes immediately.
0° SYNC
The frequency changes when the measurement signal phase
becomes 0°.
Remark
The phase of the measurement signal is the phase at the signal
source inside this instrument. It may differ from the phase of the
signal applied to the DUT.
・
Sweep
Sweep item
One of frequency, measurement signal amplitude, DC bias, and
time (zero span)
Sweep type
Either linear or log
Log can be set only when the sweep item is frequency or
measurement signal amplitude.
Sweep control
SWEEP UP:
Sweeps in the direction of lower limit → upper
limit.
SWEEP DOWN: Sweeps in the direction of upper limit → lower
limit.
SPOT:
Measures with fixed frequency, measurement
signal amplitude, and bias.
REPEAT:
Repeats SWEEP or SPOT when turns on.
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...