3.5 Setting Screen Reference
3
-
43
ZA 57630
3.5.1 Mode Menu
This menu sets the measurement mode of this instrument.
○ Setting procedure
<List operations (requiring confirmation)>
・
IMPD
−
EXT
: Measure the impedance. Current-voltage conversion needs to be
performed outside of this instrument.
The
OSC
,
PORT1
, and
PORT2
connectors are used.
・
IMPD
−
3T
: Measure the impedance. Measurement is performed using a test fixture
for an LCR meter.
The
H
CUR
, H
POT
, L
POT
,
and
L
CUR
connectors are used.
・
IMPD
−
2T
: Measure the impedance. Measure a DUT with a coaxial structure.
The
PORT3
connector is used.
・
G
−
PH
: Measure the gain and phase.
The
OSC
,
PORT1
, and
PORT2
connectors are used.
(Supplementary explanation)
When the measurement mode is changed, the frequency, amplitude, and other
measurement parameters are initialized.
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...