5.3
Measurement Connectors (Measurement Mode IMPD−EXT)
5
-
7
ZA57630
5.3.2 Front Panel H
POT
/PORT1 and L
CUR
/PORT2
・
Input connectors
BNC connectors (front panel
H
POT
/PORT1
and
L
CUR
/PORT2
)
・
Input Impedance
1 MΩ ±2%, 25 pF ±5 pF (
H
POT
) / 30 pF ±5 pF(
L
CUR
) in parallel
・
Maximum non-destructive input voltage
±20 V
・
Measurement range 10 mVrms to 5 Vrms(1−2−5 sequence) , 7 Vrms, and AUTO
PORT1 and PORT2 can be set individually.
Measurement range and maximum measurement input voltage
Measurement
range [rms]
Maximum measurement
input voltage
10 mV
±16 mV
20 mV
±31 mV
50 mV
±78 mV
100 mV
±160 mV
200 mV
±310 mV
500 mV
±780 mV
1 V
±1.6 V
2 V
±3.1 V
5 V
±7.8 V
7 V, AUTO
±11 V
・
Input weighting factor PORT1 and PORT2 can be set individually.
Setting range
±(1.00000E−15 to 999.999E+09)
Setting resolution 6 digits or 1E−15
Remark
This function corrects the conversion ratios of the voltage probe,
current probe, shunt resistance, etc. for measurement.
・
Over detection
Setting range
H
POT
/PORT1
0 to 7 Vrms
L
CUR
/PORT2
0 to 7 Vrms
Setting resolution 3 digits or 1 µVrms, whichever is the largest.
Processing
Buzzer alarm sound (can be turned on/off)
Stopping of measurement (can be turned on/off)
5.3.3 Rear Panel Normal DC Bias Output
・
Connector
BNC connector (rear panel
DC BIAS OUT
)
・
Setting range and setting resolution
(Common to "
・
Normal DC bias" in "
5.3.1 Front Panel H
CUR
/OSC
")
・
Accuracy
±(|1% of normal DC bias setting [V]| + 30 mV)
・
Output Impedance
600 Ω (nominal value)
・
Remark
(Normal DC bias setting [V] ÷ K) [V] is output from the
DC BIAS
OUT
connector.
Either the front panel
H
CUR
/OSC
or rear panel
DC BIAS OUT
is
selected for the output connector of the normal DC bias.
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...