5.4
Measurement Connectors (Measurement Mode IMPD−3T)
5
-
9
ZA57630
Setting resolution
Voltage
3 digits or 10 µVrms, whichever is the largest
Current
3 digits or 100 nArms, whichever is the largest
Accuracy
When voltage setting (Common to "
・
Measurement signal amplitude" in "
5.3.1 Front
Panel H
CUR
/OSC
")
When current setting (nominal value)
Frequency characteristics
(Common to "
・
Measurement signal amplitude" in "
5.3.1
Front Panel H
CUR
/OSC
")
Distortion
(Common to "
・
Measurement signal amplitude" in "
5.3.1 Front Panel
H
CUR
/OSC
")
ALC
{CV (constant voltage) or CC (constant current)} / OFF
Operation is with a CV when the measurement signal amplitude is
set as a voltage and with a CC when it is set as a current.
The range that can be controlled constantly will be narrower than
the setting range of the measurement signal amplitude depending
on product variations or DUT impedance.
Output limit
Setting range
Voltage
10 µVrms to 3.00 Vrms
Current
100 nArms to 60 mArms
Setting resolution
Voltage
3 digits or 10 µVrms, whichever is the largest
Current
3 digits or 100 nArms, whichever is the largest
Remark
This function prevents an excessive signal level being output due
to an incorrect setting. A measurement signal amplitude exceeding
the output limit setting cannot be output.
Remark
The setting unit (voltage or current) is common for the
measurement signal amplitude and normal DC bias.
・
Normal DC bias
Setting range
Voltage
−5.00 V to +5.00 V
There is a following restriction between the measurement signal
amplitude setting and normal DC bias setting.
(Measurement signal amplitude setting [Vrms] × 1.42) + |Normal DC bias
setting [V]| ≤ 5.0
Current
−100 mA to +100 mA
There is a following restriction between the measurement signal
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...