5.4
Measurement Connectors (Measurement Mode IMPD−3T)
5
-
8
ZA57630
5.4 Measurement Connectors (Measurement Mode IMPD
−3T)
5.4.1 Front Panel H
CUR
/OSC
An AC signal for measurement and a DC bias can be output as signals for measurement.
The AC signal for measurement is output from the front panel
H
CUR
/OSC
connector.
There are the two types of DC bias, the normal DC bias and HV DC bias, and the selected
type can be used.
The normal DC bias and HV DC bias are output from the front panel
H
CUR
/OSC
connector.
(The normal DC bias and HV DC bias cannot be output from the rear panel
DC BIAS OUT
connector.)
A signal with the AC signal for measurement and DC bias composited is output from the
front panel H
CUR
/OSC connector.
Unless otherwise specified, the measurement signal amplitude is set as a voltage and ALC is
OFF.
・
Connector
BNC connector (front panel
H
CUR
/OSC
)
・
Frequency
Setting range
10 µHz to 36 MHz (when HV DC bias is off)
1 kHz to 36 MHz (when HV DC bias is on)
Setting resolution 10 µHz
Accuracy
(Common to "
・
Frequency" in "
5.3.1 Front Panel H
CUR
/OSC
")
・
Measurement signal amplitude
Setting range
Voltage
0 to 3.00 Vrms
There is a following restriction between the measurement signal
amplitude setting and normal DC bias setting.
(Measurement signal amplitude setting [Vrms] × 1.42) + |Normal DC bias
setting [V]| ≤ 5.0
There is a following restriction between the measurement signal
amplitude setting and HV DC bias setting.
(Measurement signal amplitude setting [Vrms] × 1.42) + |HV DC bias setting
[V]| ≤ 42.0
Current
0 to 60 mArms
The setting range is limited because it is dependent on the
measurement range, etc.
There is a following restriction between the measurement signal
amplitude setting and normal DC bias setting.
(Measurement signal amplitude setting [Arms] × 71) + |Normal DC bias setting
[A] × 50| ≤ 5.0
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...