5.8 Measurement Accuracy
5
-
30
ZA57630
5.8.4 Measurement Accuracy of Measurement Parameters Other Than Z and
θ
(Measurement Modes IMPD
−EXT, IMPD−3T and IMPD
−
2T)
Calculate the measurement accuracy from the impedance measurement accuracy as follows.
Here, Q
X
is the measurement value of Q, D
X
is the measurement value of D, and θ
X
is the
measurement value of θ. It is also acceptable to calculate the θ
X
used for the accuracy
calculation by either (90° − tan
−1
|1/Q
X
|) or (90° − tan
−1
|D
X
|).
Parameter
Measurement accuracy (supplementary value)
|Y|
,
ε
S
,
µ
S
±A
Z
[%]
L
P
,
L
S
,
X
,
ε
S
'
,
µ
S
'
±A
Z
[%] (|Q
X
|≥ 10)
,
±A
Z
/ sinθ
X
[%] (|Q
X
|< 10)
C
P
,
C
S
,
B ±A
Z
[%] (|D
X
|≤ 0.1)
,
±A
Z
/ sinθ
X
[%] (|D
X
|>0.1)
R
P
,
R
S
,
G
,
ε
S
"
,
µ
S
"
±A
Z
[%] (|Q
X
|≤ 0.1)
,
±A
Z
/ cosθ
X
[%] (|Q
X
|>0.1)
Q
±Qx
2
×P
e
/ (1−|Q
X
|×P
e
)
(|Q
X
|≥ 10 or |Q
X
|×P
e
≤ 0.1)
Here, phase angle error P
e
[rad] = P
Z
[° ] / 57.3.
The measurement accuracy of Q is the actual value and not the %
value.
D
±(P
Z
[° ] / 57.3.)
(|D
X
|≤ 0.1)
The measurement accuracy of D is the actual value and not the %
value.
5.8.5 Measurement Accuracy of Measurement Parameters Other Than Gain and
θ
(Measurement Mode G
−PH)
Calculate the measurement accuracy from the phase measurement accuracy as follows. Here,
P
G
is the measurement accuracy [°] of θ.
Parameter
Measurement accuracy (supplementary value)
GD
±
P
G
360×APT [s]
Here, APT is the aperture frequency (
⊿
f [Hz]), and is aperture
setting
*1
× sweep measurement frequency interval.
*1: "Aperture setting" is a parameter that is set in this instrument for group delay (GD)
measurement.
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...