3.5 Setting Screen Reference
3
-
102
ZA 57630
○ Functions of function keys
・
[EDIT]
: Set the distance between electrodes (unit: mm) with the numerical input
palette.
・
[MODIFY]
: Modify the distance between electrodes (unit: mm) with the knob or
cursor keys.
■ [REL PERMB]:
Set parameters for relative permeability measurement.
This sets the shape information of the DUT for displaying the relative permeability of the
DUT. It can be set when the measurement mode is IMPD−EXT, IMPD−3T and IMPD−2T.
When relative permeability related (μ
S
, μ
S
', μ
S
'', or D
μ
) measurement is selected in
[Graph]
-
[DISPLAY MODE Y]
or
[Measure]
-
[PRMTR
n
], the relative permeability will be calculated
using the parameters that you set here.
■ [REL PERMB]
-
[CORE AREA]:
Set the core cross-sectional area.
○ Setting procedure
<Numerical settings (requiring confirmation)>, <Numerical settings>
○ Functions of function keys
・
[EDIT]
: Set the core cross-sectional area (unit: mm
2
) with the numerical input
palette.
・
[MODIFY]
: Modify the core cross-sectional area (unit: mm
2
) with the knob or cursor
keys.
■ [REL PERMB]
-
[MAG PATH LENGTH]:
Set the magnetic path length.
○ Setting procedure
<Numerical settings (requiring confirmation)>, <Numerical settings>
○ Functions of function keys
・
[EDIT]
: Set the magnetic path length (unit: mm) with the numerical input palette.
・
[MODIFY]
: Modify the magnetic path length (unit: mm) with the knob or cursor keys.
■ [REL PERMB]
-
[COIL TURNS]:
Set the number of coil turns.
○ Setting procedure
<Numerical settings (requiring confirmation)>, <Numerical settings>
○ Functions of function keys
・
[EDIT]
: Set the number of coil turns with the numerical input palette.
・
[MODIFY]
: Modify the number of coil turns with the knob or cursor keys.
■ [REL PERMB]
-
[WIRE DIAMETER]:
Set the winding wire diameter.
○ Setting procedure
<Numerical settings (requiring confirmation)>, <Numerical settings>
○ Functions of function keys
・
[EDIT]
: Set the coil winding wire diameter (unit: mm) with the numerical input
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...