3.5 Setting Screen Reference
3
-
75
ZA 57630
■ [MEAS TRACE COPY]
-
[OPEN_PEXT CORR MEM]
:
Copy to port extended end open correction data.
■ [MEAS TRACE COPY]
-
[SHORT_PEXT CORR MEM]
:
Copy to port extended end short correction data.
■ [MEAS TRACE COPY]
-
[LOAD_PEXT CORR MEM]
:
Copy to port extended end load correction data.
This can be set only when the measurement mode is IMPD−EXT, IMPD−3T, and IMPD−2T.
○ Setting procedure
<Execution operations (immediate execution)>
○ Functions of function keys
・
[COPY]
: Copy to the corresponding specified correction data.
The copy destination correction data numbers (1 to 32) are the numbers specified in the following.
[Calibration]
-
[OPEN CORR]
;
Open correction data
[Calibration]
-
[SHORT CORR]
;
Short correction data
[Calibration]
-
[LOAD CORR]
;
Load correction data
[Calibration]
-
[PORT EXT]
-
[OPEN_PEXT CORR];
Port extended end open correction data
[Calibration]
-
[PORT EXT]
-
[SHORT_PEXT CORR];
Port extended end short correction data
[Calibration]
-
[PORT EXT]
-
[LOAD_PEXT CORR];
Port extended end load correction data
For the equalizing, open correction, short correction, and load correction operating
procedures, refer to
“1.8 Error Correction”
in
Instruction Manual (Advanced)
.
■ [AUTO STORE]
: Set automatically copying of the measurement data to the reference data.
This is a function to automatically copy the MEAS data to the REF1 data, the original REF1
data to the REF2 data, the original REF2 data to the REF3 data, and so on every time one
sweep finishes. The last REFn data is discarded.
○ Setting procedure
<Numerical settings (immediate execution)>, <Numerical settings>
○ Functions of function keys
・
[ON] / [OFF]
: Set the automatic copy function to ON or OFF.
・
[EDIT
]
: Set the last reference data copy destination with the numerical input palette.
・
[MODIFY]
: Modify the last reference data copy destination with the knob or cursor keys.
The setting range is 1 to 8 and OFF. Specify the last REF data number to copy data. If OFF
is specified, the automatic copy function will turn off.
■ [DELETE MEAS]
: Delete the measurement data.
This deletes the measurement data (MEAS DATA) from memory. The graph traces are also
deleted. The data cannot be recovered after deletion, so save the data to a USB memory
Содержание ZA57630
Страница 1: ...NF Corporation IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS...
Страница 2: ......
Страница 3: ...IMPEDANCE ANALYZER ZA57630 INSTRUCTION MANUAL BASICS DA00068820 003...
Страница 22: ...1 8 ZA57630 Blank...
Страница 28: ...2 2 Assembly and Installation 2 6 ZA 57630 Figure 2 1 Rack Mount Kit Assembly Drawing JIS...
Страница 29: ...2 2 Assembly and Installation 2 7 ZA 57630 Figure 2 2 Rack Mount Kit Assembly Drawing EIA...
Страница 30: ...2 2 Assembly and Installation 2 8 ZA 57630 Figure 2 3 Rack Mount Dimensions JIS...
Страница 31: ...2 2 Assembly and Installation 2 9 ZA 57630 Figure 2 4 Rack Mount Dimensions EIA...
Страница 36: ...2 14 ZA 57630 Blank...
Страница 224: ...5 14 Miscellaneous Specifications 5 40 ZA57630 Figure 5 1 External dimensions...
Страница 240: ...XVI ZA57630 Blank...
Страница 242: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...http www nfcorp co jp NF Corporation 6 3 20 TsunashimaHigashi Kohoku ku Yokohama223 8508JAPAN...