16 - 58
Operator’s Manual
16 Probes and Biopsy
3.
Remove any residue with a water-moistened soft cloth on the magnetizer.
4.
Wipe off water on the magnetizer using sterile cloth or gauze after washing.
Disinfecting by Immersion
Perform the following procedure:
1.
Wear sterile gloves to prevent infection.
2.
Clean the magnetizer before disinfecting it. MINDRAY recommends the compatible solutions
to disinfect the magnetizer.
–
Refer to the instructions provided by the chemical manufacturer concerning concentration
of the disinfectant solution, method of disinfection and dilution and cautions during use.
–
Soak the magnetizer into the disinfectant solution for the shortest time the manufacturer
recommends.
–
Follow local regulations when selecting and using the disinfectant.
3.
Rinse the magnetizer with plenty of water (about 2 gallons) for at least 1 minute to remove all
chemical residues on it. Or, follow the rinsing method recommended by the disinfectant
manufacturer to rinse the magnetizer.
4.
Wipe off the water on the magnetizer with sterile cloth or gauze after rinsing it. Do not dry the
magnetizer by heating.
16.4.6 Storage and Transportation
When all examinations for the day have been completed, confirm that the magnetizer is in good
condition. After disinfecting the magnetizer, confirm that the magnetizer is in good condition and
stored in a suitable place.
•
To prevent the magnetizer from being damaged, DO NOT store it where it may be exposed to:
–
Direct sunlight
–
Sudden changes in temperature
–
Dust
–
Excessive vibration
–
Heat generators
•
Store and transport the magnetizer in the ambient temperature range of -40 °C to 50 °C.
•
Sterilize the carrying case as necessary.
16.4.7 Disposal
Be sure to disinfect the magnetizer before disposing of it.
Contact your MINDRAY representative when disposing of this device.
Содержание Ana
Страница 2: ......
Страница 50: ...This page intentionally left blank...
Страница 60: ...This page intentionally left blank...
Страница 110: ...This page intentionally left blank...
Страница 116: ...This page intentionally left blank...
Страница 166: ...This page intentionally left blank...
Страница 176: ...This page intentionally left blank...
Страница 194: ...This page intentionally left blank...
Страница 220: ...This page intentionally left blank...
Страница 288: ...This page intentionally left blank...
Страница 304: ...This page intentionally left blank...
Страница 308: ...This page intentionally left blank...
Страница 316: ...This page intentionally left blank...
Страница 337: ......
Страница 338: ...P N 046 018835 00 2 0...