4 - 12

Operator’s Manual

  4  Setup

4.1.9 Output

Set the output format, the range and the resolution for the image.

4.1.10 Access Control

The system supports two types of users: administrator and operator.

Administrator
The system administrator can access all function modules, and view all patient data, such as 
patient information, images and reports, etc. Only one administrator is configured by default. 
The administrator can add or delete operators. 

Operator
The operator can only access the function modules with assigned privileges. The operator can 
only view exam information saved in the system and operated by him or herself, such as 
patient information, images and reports, etc.

Enabling Access Control

The system administrator can preset the access controls, that is, whether an operator has the right to 
access data in the system.
Access control only can be set by the system administrator.
Open the “Access Control” page:

If “Enable User Account Control” is selected, you must be authorized before accessing the 
data, and you can configure password policy and LDAP, and change password. If unselected, 
you can access all the data without authorization, and you cannot configure password policy 
and LDAP, and change password.

If “Enable Emergency User” is selected, the administrator can edit privileges for emergency 
users. If unselected, the administrator cannot edit privileges for emergency users. 

Type

Item

Description

AVI ENCODE

Encode Quality

To set the image quality of unloaded AVI. The system 
unloads according to the settings.
The higher the image quality is, the clearer the 
unloaded image is. The unloading speed become 
slower with the larger space.

Operating System 
Compatibility

To set the unloading format of the AVI. The system 
unloads according to the settings.
If checking “Mac OS”, saving CIN files to USB flash 
drive as “MP4 Video”.

Send To Frame Rate To set whether to enable/disenable compression of 

images.

Analog Output 
Mode

To set the format to output/separate the video format: 
NTSC or PAL.

DIGITAL 
OUTPUT

Output Size

Select VGA, HDMI output image range.

Resolution

Select VGA, HDMI output image resolution.

DVR OUTPUT

Output Size

Select DVR recording area.

Содержание Ana

Страница 1: ...us7 Zeus ME7 ME7 Exp Anesus ME7 Anesus ME7 Pro Anesus ME7S Anesus ME7T Ana Anesus Vaus Anesus Zeus Crius ME7P Crius ME7W Anesus ME7P Anesus ME7W Emerus ME7P Emerus ME7W MX7P MX7W Diagnostic Ultrasound...

Страница 2: ......

Страница 3: ...countries All other trademarks that appear in this manual are used only for informational or editorial purposes They are the property of their respective owners Responsibility on the Manufacturer Part...

Страница 4: ...Malfunction or damage caused by improper operation or repair by unqualified or unauthorized service people Malfunction of the instrument or part whose serial number is not legible enough Others not ca...

Страница 5: ...u shall also be continuously alert to dangers other than those indicated Mindray shall not be liable for damage or loss resulting from negligence or ignorance of the precautions and operating instruct...

Страница 6: ...ettings and optional configuration the actual interfaces may be different from those in this manual Conventions In this manual the following conventions are used to describe the buttons on the control...

Страница 7: ...ety Classifications 2 1 2 4 Product Specifications 2 2 2 4 1 Imaging Mode 2 2 2 4 2 Power supply 2 2 2 4 3 Environmental Conditions 2 2 2 4 4 Dimensions and Weight 2 3 2 5 Product Differences 2 3 2 6...

Страница 8: ...r Brightness Contrast Adjustment 3 5 3 5 Connecting Disconnecting a Probe 3 6 3 5 1 Connecting a Probe 3 6 3 5 2 Disconnecting a probe 3 7 3 6 Connecting USB Devices 3 7 3 7 Connecting the Footswitch...

Страница 9: ...eGateway Preset 4 43 4 9 Print Preset 4 44 4 9 1 Print Setting 4 45 4 9 2 Image Settings 4 45 4 10 Maintenance 4 45 4 10 1 Option 4 45 4 10 2 Exporting Setup Data 4 46 4 10 3 Importing Setup Data 4 46...

Страница 10: ...6 8 1 PW CW Mode Image Scanning 6 14 6 8 2 PW CW Mode Image Parameter 6 14 6 9 iScape 6 17 6 9 1 Basic Procedures for iScape Imaging 6 17 6 9 2 Image Review 6 18 6 9 3 Evaluate image quality 6 18 6 9...

Страница 11: ...ging 9 2 9 4 Image Parameters 9 2 9 5 Image Saving 9 4 9 6 Micro Flow Enhancement 9 4 9 7 Contrast Imaging QA 9 4 10 Physiological Unit Signal 10 1 10 1 ECG 10 1 10 2 Respiratory Wave 10 3 10 3 ECG Re...

Страница 12: ...2 13 2 1 Adding Comments 13 3 13 2 2 Moving Comments 13 4 13 2 3 Editing Comments 13 4 13 2 4 Deleting Comments 13 4 13 2 5 Setting the Start Position 13 5 13 3 Body Mark 13 5 13 3 1 Adding Body Mark...

Страница 13: ...ter sending images on the iStation screen 15 5 15 5 2 To send storage commitment automatically after an exam ends 15 5 15 6 Query Retrieve 15 5 15 7 DICOM Media Storage DICOMDIR Review 15 6 15 7 1 Med...

Страница 14: ...18 1 3 Cleaning the peripherals 18 6 18 1 4 Checking the Probe 18 6 18 1 5 Checking the Power Cable and Plug 18 7 18 1 6 Checking Appearance 18 7 18 1 7 System Hard Drive Backup 18 7 18 2 Troubleshoot...

Страница 15: ...2 D 3 EAP Network D 3 E iScanHelper E 1 E 1 Use iScanHelper for Reference E 1 E 2 Use iScanHelper for Learning or Training E 1 E 3 Basic Screen and Operation E 2 E 3 1 View Selecting Area E 2 E 3 2 S...

Страница 16: ...c Output Parameters H 4 H 7 2 Limits of Acoustic Output H 5 H 7 3 Differences between Actual and Displayed MI and TI H 5 H 8 Measurement Uncertainty H 6 H 9 References for Acoustic Power and Safety H...

Страница 17: ...ic shock may result Do connect the power plug of this system to wall receptacles that meet the ratings indicated on the rating nameplate If adapters or multi functional receptacles are used it may cau...

Страница 18: ...may result or a gas explosion may occur Before cleaning the system disconnect the power cord from the outlet System failure and electric shock may result This system is not water proof designed Do Not...

Страница 19: ...your vendor Prolonged and repeated use of keyboards may result in hand or arm nerve disorders for some individuals Observe the local safety or health regulations concerning the use of keyboards When u...

Страница 20: ...iod of time required for the purpose of diagnosis Do not use the system to examine a fetus for a long period of time Except accessories that have been stated as sterile the system and its accessories...

Страница 21: ...tions with high humidity Turn ON the system only after the power has been turned OFF for a while If the system is turned ON immediately after being turned OFF the system may not be rebooted properly a...

Страница 22: ...each examination A defective probe may cause electric shock to the patient Do not subject the probe to shock A defective probe may cause electric shock to the patient Do not disassemble the probe to a...

Страница 23: ...t of improper gel or cleaner DO NOT dip the transducer in the strong polar solution of ethanol chloride of lime ammonium chloride acetone and formaldehyde DO NOT contact the transducer with solution o...

Страница 24: ...EE label applies to EU member states only The use of this symbol indicates that this system should not be treated as household waste By ensuring that this system is disposed of correctly you will help...

Страница 25: ...s rated as IPX8 The power adapter is rated as IP22 According to the disinfection and sterilization method s recommended by manufacturer Equipment with disinfection and sterilization method s recommend...

Страница 26: ...tery voltage 14 4 V 2 4 3 Environmental Conditions Operating conditions Ambient temperature 0 C 40 C Relative humidity 20 85 no condensation Atmospheric pressure 700 hPa 1060 hPa Storage and transport...

Страница 27: ...ions and Weight Dimensions D W H unit mm 322 5 364 5 44 3 Weight including batteries 4 0 kg 2 5 Product Differences Model Histogram Profile Color Vel Double Dist Spline length Parallel line MX7 MX7 Ex...

Страница 28: ...defined label Cables 2 6 2 Probes and Needle guided Brackets Available Please see 16 Probes and Biopsy 2 6 3 Options MX7W Model Histogram Profile Color Vel Double Dist Spline length Parallel line Item...

Страница 29: ...t Cardiology Package should be configured Strain Elastography Stress Echo Cardiology Package should be configured Tissue Tracking QA Cardiology Package should be configured Smart 3D DICOM DICOM Basic...

Страница 30: ...erve Package Emergency Critical Package RIMT Vascular Package should be configured AutoEF Cardiology Package should be configured R VQS Vascular Package should be configured Smart B line Smart VTI Car...

Страница 31: ...system is connected to Class B peripheral devices otherwise RF interference may occur and the customer or user must take adequate measures accordingly 2 6 5 Parts that can be used within patient envi...

Страница 32: ...on Power indicator Used for turning on off the power The indicator lights up when the system is powered on 4 S Video S Video signal output 5 Network port Connects to network 6 HDMI Used for extending...

Страница 33: ...the actual buttons and knobs 11 Probe port Connects a probe to the main unit or connects a probe extend module 12 Probe locking lever Locks or unlocks the probe connected with the main unit locked sym...

Страница 34: ...types of screws After fixing the module using handle screws fold the handles parallel to the bottom of the main unit 1 Insert the connector to the probe port of the main unit 2 Toggle the locking lev...

Страница 35: ...he U Bank installation and uninstallation If you need to change the batteries in the U Bank please contact your sales representative Specifications are as follows Voltage 19V DC Capacity 26400 mAh or...

Страница 36: ...blinks remaining capacity 20 5 Baterry status indicator Indicates the charge discharge status of the U Bank When the battery is charging Not fully charged the indicator blinks in orange Fully charged...

Страница 37: ...re above until the system is locked in place TIP To remove the U Bank from the system press the detach lever of U Bank and lift up slightly the system by holding the system s handle 2 10 ECG Module Ov...

Страница 38: ...gured with trackpad 2 ECG lead port Used for ECG signal input No Name Description No Name English Name Description 1 TDI TDI mode button Enters TDI mode P4 User defined key Undefined button set by the...

Страница 39: ...ual split window button Press to enter the Dual mode from another mode Press to switch between the two windows in the Dual mode 12 Image zoom Press to enter or exit zoom status 13 iTouch Gain Gain and...

Страница 40: ...lluminates in green color when the power of the batteries is sufficient It illuminates in orange color for low battery power 21 Clear Clear button Press to clear off the comments or measurement calipe...

Страница 41: ...ion screen 3 Probe exam mode switch button Tap to switch probe and exam mode 4 Scan button Tap to enter B imaging mode 5 Patient data management system button Tap to enter or exit patient information...

Страница 42: ...ment calipers color bar grayscale bar are also displayed here 4 Hint trackball trackpad and Set key function indications Displays the current functions of trackball trackpad and Set keys 5 User define...

Страница 43: ...ation and close the screen To reposition a dialog box 1 Use the trackball trackpad to move the cursor onto the title bar of the dialog box At this time the cursor becomes a then press Set key 2 Use th...

Страница 44: ...Parameter magnitude setting Tap the left right part of the parameter button or tap to select the parameter button and rotate the multi functional knob to increase decrease the value ON OFF setting so...

Страница 45: ...ter bodymark status 3 Comment tab Tap to enter text arrow comment status 4 Page If there is more than one page of comment texts for the current exam mode you can slide to view more 5 Comment position...

Страница 46: ...Panel 12 Language button 13 Arrow button Tap to add an arrow comment to the image 14 Library Tap to pop up and select the library No Item Description Function Operations Adding a control Tap to select...

Страница 47: ...microphone the system recognizes the vocal order The system conducts the operations after recognizing the voice 2 Click to stop the voice Recognition 3 Click to open the help information No Warning La...

Страница 48: ...ator Battery indicator When the lever located at the bottom of the monitor support arm points to you can move the monitor to the right and left When the lever located at the bottom of the monitor supp...

Страница 49: ...roduct is provided with a CE marking in accordance with the regulations stated in Council Directive 93 42 EEC concerning Medical Devices The number adjacent to the CE marking 0123 is the number of the...

Страница 50: ...This page intentionally left blank...

Страница 51: ...his system can work normally only when it is connected to the external power supply or the battery capacity is sufficient 3 2 1 Connecting Power 1 Connect the connector of the power adapter to the ada...

Страница 52: ...permanently damaged Battery specification Voltage 14 4 V Capacity 6600 mAh single battery Battery Status Indicator The battery status indicator is located in the bottom right corner of the screen ind...

Страница 53: ...e clean The locking mechanism of the casters if there is a trolley works normally 3 3 2 Power the System ON Press the power button to power the system on If access control has been set by the system a...

Страница 54: ...inuous absent or black items in the B mode image Check whether there is abnormal heat on the surface of the probe during an ultrasound procedure If you use a probe which is giving off excessive heat i...

Страница 55: ...e the batteries will be out of power and permanently damaged If the system will not be used for a long period of time you should disconnect the power adapter disconnect the mains power and turn off th...

Страница 56: ...m falling off or becoming damaged Only use probes provided by Mindray Aftermarket probes may result in damage or cause a fire 3 5 1 Connecting a Probe WARNING The probes cables and connectors should b...

Страница 57: ...sfully and the symbol appears in the bottom right corner of the screen To remove the USB device click to open the Remove USB Device screen Select the device to be removed and click OK A sound is heard...

Страница 58: ...ord and data cable The power cord shall be directly connected to a wall receptacle as required Perform the following procedure 1 Connect the data cable to USB port of the ultrasound device 2 Power on...

Страница 59: ...rget printer from the drop down list in the Property box and set other printing properties f Click OK to complete 3 8 3 Connecting a Wireless Printer The system supports the wireless graph text printe...

Страница 60: ...This page intentionally left blank...

Страница 61: ...en To exit Setup Select OK in the Setup menu The parameter settings are saved Select Cancel in the Setup menu to close the Setup menu When you change the system language and click OK in the Setup menu...

Страница 62: ...s and 8 24 32 bitdepth Time Zone To select the time zone Time Format To select the time format Date Format To set the date format System Date To set the date for the system System Time Move the cursor...

Страница 63: ...he DICOM server for storage print Send SR after End Exam Select whether to automatically send structure report to the DICOM server Maximum number of exams to be kept To set the maximum number of exams...

Страница 64: ...settings back to factory IVOCAL Automatically shut down waiting time To set the delay before voice recording stops automatically Enable Voice Response After it is ticked the system will repeat record...

Страница 65: ...e imaging mode B Color Refresh with PW CW Sampling Line Movement To set whether to turn on the function that when moving PW CW sampling line B Color image is activated under B Color PW CW mode Uninter...

Страница 66: ...ze of the cursor Heart Beat The number of cardiac cycles in the heart rate calculation In heart rate measurement the number of cardiac cycles should match the preset number Cursor Line Display If unse...

Страница 67: ...Fetal Growth or Fetal Weight page select an OB Item in the left column 2 Select a formula in the right column 3 Click Default The default formula is marked with a To set the fetal weight display Perf...

Страница 68: ...95 10 90 Unit of the measurement value according to the table to import select from mm cm g kg cm or mm Row number N of the table the maximum row number N in the column No The third row is empty GA va...

Страница 69: ...osis You can add user defined formulae for items obstetric tools that are not included in the GA and FG table 1 Select More OB Items on the Fetal Gestational Age or Fetal Growth page 2 Select an item...

Страница 70: ...e steps above Trackball Speed Set the speed of the trackball when using the trackball Key Volume Set the key volume level Key Brightness Set the brightness for the keys Trackball Light Set the light c...

Страница 71: ...ow the available auxiliary functions for current key Select Send Image to DICOM Storage in the Output column on the right side Then the Function 2 on the left side turns into Send Image to DICOM Stora...

Страница 72: ...horized before accessing the data and you can configure password policy and LDAP and change password If unselected you can access all the data without authorization and you cannot configure password p...

Страница 73: ...ed user Editing Privilege Turn on the access control function and log in to the system as Administrator before you edit privileges 1 Select a user click Edit Privilege to enter the Edit user privilege...

Страница 74: ...t the duration after an account is locked Reset all lockout Reset all locked accounts Enable strong password Enable strong password to improve security If the strong password is enabled and you log in...

Страница 75: ...rs and privileges will appear in the Member of filter list Select a member to be deleted and click Delete Select a member to be modified modify the member name and select or deselect privileges from t...

Страница 76: ...set screen press Cursor to show the cursor and move the cursor to the bottom right corner of the screen Click icon to see the scene management list b Click Configure to see the following dialog box c...

Страница 77: ...n Operator or Diagnostician item in Patient page automatically After scanning a 1D barcode of an operator or diagnostician for a second time the obtained data is B and A will be appended by B in Opera...

Страница 78: ...y service engineer Item Description Item Description Scan Barcode Example Input a barcode example barcode example is separated by separators the separator is used to set the start and end position of...

Страница 79: ...add all exam modes in the library to the Probe and Exam Mode list Click add all probe and exam modes in the library to the Exam Mode Library list Click Delete to delete a user defined exam in the Exa...

Страница 80: ...Select the 2D M or Doppler tab to go to the corresponding preset menu Available Items general measurement tools configured by the system in the current scanning mode which are available but not assign...

Страница 81: ...tion Item Name Result Results obtained from D trace are listed The selected items will be displayed in the results window after measurement If PV is selected other results become deselected except the...

Страница 82: ...ting Package Positions Creating Packages Click New and enter a name for the new package in the dialog box pop up Click OK to confirm the new package displays in the Available Items list Adding Removin...

Страница 83: ...gonometric functions are in degrees not radians PI is accurate to 7 digits Adding B Hist or B Profile to the study is not supported Click Export Custom in the measurement preset window to export the u...

Страница 84: ...a Select User Defined in the Available Items and select the desired item b Click Delete Remove Study Select a user defined study click 4 3 3 Report Preset NOTE Deleting is not supported in IVF IMT and...

Страница 85: ...g and exit 7 Change the patient information layout in the report template Change the template used in the report layout click New Layout to select another template Double click the information lines t...

Страница 86: ...displayed below Available Items 2 Click and to add the item in Available Items on the left into Selected Items on the right TIP You can only delete the user defined items rather than the items in the...

Страница 87: ...can customize the protocols and views in the iWorks preset screen 4 5 1 Protocol Management Click to select the protocol in the list The protocol type can be checked on the right Click Add Protocol to...

Страница 88: ...tocol Or select an existing protocol and click Copy to customize the protocol based on the previous template 2 Enter the protocol name type and select the application region 3 Click Add View to enter...

Страница 89: ...e list b Enter a phase name Select Auto Select the system jumps to Select Mode after retrospective acquisition Item Description Protocol Name Enter the protocol name Trigger Set the trigger type WMS m...

Страница 90: ...protocol Deleting a Stress Echo Protocol Click an user defined protocol on the Select Protocol screen click Delete Protocol 4 6 2 Maintenance Item Description Acquire mode Set the type of ROI manual R...

Страница 91: ...value is less than 16384 or greater than 65536 the system automatically sets it to the value 32768 DICOM output charasets Select an character set for DICOM output according to the local PACS workstat...

Страница 92: ...e from the Strategy List and assign strategy items to the selected strategy Add Set the function from the drop list box enter the parameter 1 and parameter 2 and click Add Then the added strategy item...

Страница 93: ...correct application name Storage Service Preset DICOM storage preset items are described as follows Item Description Device After you set the server s in DICOM Preset screen the name s will appear in...

Страница 94: ...when choosing the mix The image use 8 bit if the image is captured in color mode or the image has the tint All images use 8 bit when choosing the grey Allow Multiframe If SCP supports this function th...

Страница 95: ...nd white printing Please select the type the printer supports Film Orientation Select between LANDSCAPE and PORTRAIT Priority Specify printing task priority among HIGH MED and LOW Film Size Select fil...

Страница 96: ...the service TLS Transport Layer Security Select whether to encrypt the data during network transportation Strategy Name Set the DICOM strategy Remove Attributes 0 Set what DICOM element s that will no...

Страница 97: ...me s Reserved feature Timeout Refers to the amount of time after which the system will stop trying to establish a connection to the service Associated Storage Service The associated storage server is...

Страница 98: ...y title here it should be consistent with that of the HL7 server Port DICOM communication port 104 by default Here the port should be consistent with that of the HL7 server port Maximum Retries Reserv...

Страница 99: ...N is successfully connected Disconnected VPN is disconnected Error error connection Server IP Group User Name Password Hide characters The password is displayed as Connect Disconnect Connect or discon...

Страница 100: ...vice to be updated in the service list 2 You can see properties in the Configure Service area 3 Do one of the following to edit the network service Modify the properties and click Update to update the...

Страница 101: ...twork website service you can check data using the website authorized account and password provided by the service vendor You can open the browser Q View to review previously sent DICOM data Q path is...

Страница 102: ...rver Password On Worksheet Sets whether to display the Signature field box in a worksheet Tap Report WorkSheet or tap Review Report WorkSheet enter the worksheet password in the field box and click OK...

Страница 103: ...eway service 1 Set the eGateway server properties as described above 2 Tap Add to add the service to the service list To modify a network service 1 Select the service to be updated in the service list...

Страница 104: ...t Port for transmitting DocServer IP Address IP address of the eGateway service Port Port for transmitting Image Cine PDF Report Set the shared document content Clear Click to cancel the parameter set...

Страница 105: ...ire other maintenance functions please contact Mindray Customer Service Department or sales representative 4 10 1 Option The system enters the Option page after entering the Maintenance screen In the...

Страница 106: ...em The system will reset and operate according to the setup preferences that were imported Perform the following procedure 1 Click Import to open the Load Data screen 2 Select the imported file 3 Clic...

Страница 107: ...eSpacial Navi function NOTE The list of needles must be release version provided by Mindray Needles List Recover Restore the list of needles that support eSpacial Navi function to the last version Ne...

Страница 108: ...b Click Wipe and operate according to the screen prompts to clear patient data c Select User Define again and click Confirm d Input the password and click Confirm to finish the password setting NOTE...

Страница 109: ...exe or any other abnormal operation during the installation please contact the Mindray service engineers Transmission Encryption After accessing the network click VPN Config to enter the VPN Config in...

Страница 110: ...This page intentionally left blank...

Страница 111: ...a new patient exam it is better to type the detailed patient information The system will set up a unique information database for each patient based on the patient information entered so that the inf...

Страница 112: ...al causes or system power off Cancel Exam to cancel the current exam The cancelled exam can t be restored Click Quick Register to save the patient information quickly and return to the main screen OK...

Страница 113: ...the new exam simultaneously After editing the patient information in the Patient Info screen select OK to start a new exam Review Image Click to enter the Review screen Patient Infor Click to enter th...

Страница 114: ...ted exam you can modify patient ID 5 2 2 Continue an Exam In iStation screen select an exam record paused within 24 hours click Resume Exam from the menu popped up to continue the exam If you want to...

Страница 115: ...on to avoid mixing data of the next new patient To end an exam do one of the following Tap End on the touch screen to finish the current exam Click New Patient on the Patient Info screen to end the la...

Страница 116: ...This page intentionally left blank...

Страница 117: ...B to return B mode Adjust parameters to optimize the image 6 1 2 B mode Image Parameters Image Quality Used for switching B THI and adjusting the frequency The real time value of frequency is displaye...

Страница 118: ...is available only for the convex and phased probes When the scan range is adjusted to the widest the FOV position cannot be changed Steer To steer the beam the probe transmits TIP Steer is available...

Страница 119: ...image Tint Map This function provides an imaging process based on color difference rather than gray distinction TSI The TSI function is used to optimize the image by selecting acoustic speed according...

Страница 120: ...ates the flow towards the probe while the color below the color bar indicates the flow away from the probe the brighter the color is the faster the flow speed becomes while the darker the color is the...

Страница 121: ...et Size This function is an indication of the ability to detect flow which is used to adjust the accuracy of color flow The higher the sensitivity is the more sensitive indication for low velocity flo...

Страница 122: ...evels of the flow display to display the grayscale signal or color signal The color image is preferred with higher value while grayscale signals are displayed with the lower value Velocity Tag This fu...

Страница 123: ...ow velocity flows The Directional Power maps provide information of flow direction Dynamic Range This function is to adjust the transformation of echo intensity into color signal Increasing dynamic ra...

Страница 124: ...W and Power See the relevant sections for details The following introduces the specific items in TDI mode Tissue State This function is used for fast image optimization 6 4 3 TDI Quantitative Analysis...

Страница 125: ...the trackball trackpad the images in TDI review window and 2D review window are reviewed synchronously since the two images are frozen at the same time ROI movement is linked between the TDI Tissue D...

Страница 126: ...lete the drawing Press Clear to cancel the active ROI 4 Select the curve Tap Speed Strain or Strain Rate For Strain or Strain Rate tap Strain Dist on the touch screen to select the corresponding value...

Страница 127: ...de image the real time gain value is displayed in the image parameter area Increasing the gain will brighten the image and you can see more received signals However noise may also be increased Display...

Страница 128: ...determine the size and position of the color flow displayed in the color M mode image 6 7 Anatomical M Mode For an image in the traditional M mode the M mark line goes along the beams transmitted fro...

Страница 129: ...Press B to return to real time B mode 6 7 2 Free Xros CM Curved Anatomical M Mode In Free Xros CM mode the distance time curve is generated from the sample line manually depicted anywhere on the imag...

Страница 130: ...time while the vertical axis represents Doppler frequency shift PW mode provides a function to examine flow at one specific site for its velocity direction and features while CW mode proves to be much...

Страница 131: ...e parameters as per the current tissue characteristics for a better image effect Auto Calculation This function is used to trace the spectrum and calculate the PW CW mode image parameters The results...

Страница 132: ...Doppler scale or when the selected anatomical site is too deep for the selective PW Doppler scale Baseline Refers to the area where the velocity in zero in the spectrum The map changes after being ed...

Страница 133: ...o become fully proficient Therefore the measurement results can be inaccurate Exercise caution when you perform measurements in iScape mode A smooth and even speed will help produce optimal image resu...

Страница 134: ...capture Tap Stop Capture on the touch screen Press Update Wait until the acquisition completes automatically After the acquisition is completed the panoramic image will be displayed and the system en...

Страница 135: ...on arterial stiffness tracks movements of the upper and lower vessel walls and measures vessel diameter displacement coefficient of hardness and PWV dimensionless pulse wave velocity Hardness coeffici...

Страница 136: ...ame Tap Accept Result to update the result window data to the report Save the single frame and multi frame image if necessary Tap Cancel Result to recalculate RIMT and perform step 3 5 if necessary 7...

Страница 137: ...Scanning Areas to select different zone combinations for examination 4 Select a desired zone and tap Auto Calc The system automatically starts tracing the B line sampling area and automatically recog...

Страница 138: ...b to select items 6 Press the Save key to save the single frame image and B line calculation results If necessary press the Freeze key again to unfreeze the image Repeat steps 4 6 to finish calculatin...

Страница 139: ...tting the starting point for measurement and Update to switch the active end and fixed end of the caliper d Press Set to set the end point For more information about measurement see Advanced Volume 4...

Страница 140: ...s can be saved and displayed 6 13 Smart IVC Smart IVC is used to automatically measure the IVC inner diameter and calculate the change rate in the B mode image Measure the IVC Inferior Vena Cava in re...

Страница 141: ...ll trackpad to adjust the sampling line position c Press the Update key to start calculating IVC again 5 Press the Freeze key to freeze the image and finish calculating IVC The calculation results and...

Страница 142: ...screen to select left or right carotid 3 Use the trackball trackpad to locate ROI over the target area The dotted line of the ROI is in the middle of the blood vessel Press Set to confirm the positio...

Страница 143: ...arching method to trace each spot and calculate the movement so as to determine myocardial motion in a more quantitative way Tissue tracking quantitative analysis is an option NOTE Only use the probes...

Страница 144: ...nner dots of the curve Move the cursor to the exact boundary position and press Set again to set the right place Tap Start Tracking to start tracking again 6 Tap Accept Compute to calculate and displa...

Страница 145: ...no longer displays You can get the current X Y axis value by moving the cursor onto one point on the curve and if you press Set at this time the frame marker will move to the spot The segment boundar...

Страница 146: ...3 point method As shown in the following figure after operation by pressing Set to place 3 points on the image the system generates the trace automatically Manual trace method Press Set and move the c...

Страница 147: ...base section short axis view of mitral valve PSAX M short axis view of base section short axis view of papillary muscle PSAX AP short axis view of apex Parameter Adjustment Thickness adjusts the traci...

Страница 148: ...he X axis represents time s the Y axis represents rotation of the tissue Deg Circumferential Rotation Rate curve The X axis represents time s the Y axis represents rotation by time Deg s Torsion Torsi...

Страница 149: ...Tap iWorks on the touch screen or press the user defined key for iWorks to enter the protocol selection screen and tap the corresponding protocol button to enter the status After the system enters th...

Страница 150: ...automated protocols of OB the touch screen displays as following 6 16 2 View Operation In iWorks status you can perform view selection repeat replacement and delete operations using the touch screen...

Страница 151: ...xamination Suspend Exit the protocol so the user can run the system manually This is used when an unusual or atypical workflow is required You can run the system manually if necessary Start manual exa...

Страница 152: ...e M mark line always appears after being set when entering PW M TVM mode for once Press PW M to enter the corresponding mode one time This function is unavailable for frozen dual probe mode 5 To save...

Страница 153: ...However 3D imaging can obtain the reference information by overall observation 7 1 1 Terms Volume data to obtain the data collection of three dimensional object via the sequence reconstruction to two...

Страница 154: ...truction Set ROI on the 2D image with the largest section area of the fetal face Set ROI a little larger than the fetal head Perform the following procedure 1 Press the right Set key to toggle among s...

Страница 155: ...D image rendering mode Display the minimum echo intensity in the observation direction This is helpful for viewing vessels and hollow structures X Ray Set X ray as 3D image rendering mode Display the...

Страница 156: ...provided for reference only not for confirming a diagnosis Please use caution to avoid misdiagnosis 7 1 5 Wire cage When you view a 3D image on the display monitor it s sometimes difficult to recogni...

Страница 157: ...bilical cord The fetus keeps still If there is a fetal movement you need a rescan when the fetus is still Angle of a B tangent plane The optimum tangent plane to the fetal face 3D imaging is the sagit...

Страница 158: ...The arrow in the figure below indicates the movement of the probe You can move the probe in the opposite direction to the arrow 7 2 1 Smart 3D Image Acquisition NOTE In Smart 3D image scanning if the...

Страница 159: ...ode 7 2 2 Smart 3D Image Viewing Activate MPR Select Ref Image A B C or VR to activate sectional plane image MPR or 3D image VR MPR Viewing In actual display different colors of the window box and the...

Страница 160: ...e system supports the observation of 3D image from 6 directions Tap Up Down Left Right or Front Back to select the direction of the figure a c and e Tap Flip on the first page to observe by the conver...

Страница 161: ...viewing status you can render the image by adjusting the relevant parameters Tap VR or MPR on the touch screen to select the VR parameters or MPR parameters When VR button is highlighted adjust param...

Страница 162: ...The color of image change according the tint value Quick Rotation Rotate the 3D image quickly iClear Use the 3D self adapation filter to strengthen the space continuity of the tissue structure and pr...

Страница 163: ...Use the trackball trackpad to view to a certain position tap Set End d Set the auto rotation increment rotate the knob under the Increment e Tap Off to start auto rotation Use Speed to adjust the rota...

Страница 164: ...Set to trace the region When the start point and the end point coincide the region is selected or you can press Set twice to finish tracing 3 Move the cursor to the region you want to edit and press S...

Страница 165: ...rientation and press Set to start drawing press Set when ends are necessary press Set twice to finish drawing 3 Move the cursor to the region you want to edit and press Set again to edit Soft Eraser A...

Страница 166: ...This page intentionally left blank...

Страница 167: ...ap Elasto on the touch screen or press the user defined key for Elasto to enter the elastography mode The system displays two dual B E windows in real time The left one is 2D image and the right one i...

Страница 168: ...down display the elasto image appears above and the real time ultrasound image appears below Full The elasto image only displayed Map Select different maps for observation Strain mode Affect the displ...

Страница 169: ...st imaging is an option Select an appropriate probe for Contrast Imaging see 16 1 Probes 9 1 Basic Procedures for Contrast Imaging Perform the following procedure 1 Select an appropriate probe and per...

Страница 170: ...nces echo reflection by using contrast agent to perform myocardial analysis based on echocardiography technique Perform the following procedure 1 Acquire ECG signal 2 Select an appropriate probe and c...

Страница 171: ...time via the touch screen Dual Live In live mode or freeze mode set touch screen item Dual Live as ON to enable dual live function Both the contrast mode and tissue mode are displayed The THI and B im...

Страница 172: ...Perform the following procedure 1 During real time scanning tap MFE on the touch screen to start MFE imaging 2 Use MFE Period on the touch screen to select different imaging period suitable for curre...

Страница 173: ...ifferent color and its corresponding trace data is plotted using that same color 1 Contrast cineloop window Sample area indicates sampling position of the analysis curve The sample area is color coded...

Страница 174: ...e for time intensity curve and perform data analysis on time intensity curve for data table Tap Raw Curve to hide display raw curve when the button is highlighted in green raw curve is displayed Tap T...

Страница 175: ...Select the drive and enter the file name in the displayed window c Select OK to save the data and return to the QA Analysis screen All displayed ROI traces are saved in the exported file The parameter...

Страница 176: ...This page intentionally left blank...

Страница 177: ...place the ECG electrodes directly in contact the patient s heart otherwise it may lead to stop of the patient s heartbeat Do not apply the ECG electrodes if the voltage exceeds 15 volts This could pro...

Страница 178: ...ime triggering delay time and image display format 6 Freeze the triggering image and the curve and then review them 7 Tap Physio or press the user defined key for Physio to exit ECG mode and remove EC...

Страница 179: ...Real Trigger Tap Real Trigger to enable or disable the real trigger function After the Real Trigger is enabled two images are displayed respectively in two windows One is triggered by ECG and the oth...

Страница 180: ...ime 10 4 Parameter Description The physio parameters are described as follows Parameter Description ECG Source Select ECG source Gain Set the amplitude of the trace Position Set the vertical position...

Страница 181: ...s stages are limited by time rather than a maximum number of loops the system stops acquisition after two minutes Acquisition is prospective when you select the stage label and then press Save on the...

Страница 182: ...m displays a percentage marker below the selected stage indicating the progress of the continuous capture To halt saving acquired images for the selected stage tap Pause on the touch screen or press F...

Страница 183: ...w and zoom in the clip to the full screen Double click a clip The clip will be magnified Tap First Last or Previous Next on the touch screen to display another loop in the current view 3 Select the st...

Страница 184: ...ect Display Selected on the touch screen The system displays the selected phases for each selected view side by side To display all views for a specific phase Select the phase label for example Rest T...

Страница 185: ...and ASE 17 are supported In addition each segment has 3 kinds of scoring method 4 5 and 7 points Select through the Scoring control on the touch screen To assign a wall motion score WMS 1 Select a col...

Страница 186: ...am 11 5 Exiting the Stress Echo Feature To exit the Stress Echo feature Tap End or End SE exam on the touch screen 11 6 Measurement and Report Suspend the stress echo exam by selecting Suspend Exam Pr...

Страница 187: ...nsity to occur at a different location in the acoustic filed As a result the MI may change 12 2 1 Spot Zoom Perform the following procedure 1 Press Zoom to enter the sample volume definition status 2...

Страница 188: ...key for Quad to switch among the windows press B to exit splitting display mode and enter the image of the currently activated window in full screen In frozen mode the system supports imaging mode swi...

Страница 189: ...ckball trackpad the review sequence is reversed to the image storing sequence thus the images are displayed in descending order Whereas if you move the playback mark to the right by using the trackbal...

Страница 190: ...en to set the review speed The system plays the auto review region automatically 4 Tap Jump to First Jump to Last to review the first or last image 12 4 3 Cine Review in M PW CW TVD Mode Enter cine re...

Страница 191: ...Compare in Review Mode TIP For B B COLOR B TVI B POWER B TEI mode image you can select at most 4 images for PW M CW TVD mode image you can select at most 2 images Perform the following procedure 1 Tap...

Страница 192: ...age windows cine replaying can t be performed for single frame image file press Update or Dual key to switch the active image window 3 Save the image if it is necessary Measurements adding comments an...

Страница 193: ...choose the type of the saving time and the cardiac cycles Use Time Post or Cycle Post to adjust it Prospective Cine Length Retrospective cine duration set the time that user taps Retro when playing th...

Страница 194: ...This page intentionally left blank...

Страница 195: ...otherwise false readings and potential misdiagnosis may result CAUTION If an image is unfrozen or the mode is changed during a measurement the calipers and measurement data will be cleared from the s...

Страница 196: ...or add the trace For products equipped with the trackpad set user defined keys for comments text arrow and trace first see 4 1 7 Key Configuration WARNING You must ensure that the entered comments ar...

Страница 197: ...ow button or press the user defined key for Arrow to enter the arrow status 2 Adjust the position and orientation of the arrow Use the trackball trackpad to move the arrow to the desired position Rota...

Страница 198: ...to move the cursor to the desired location to insert delete characters you can either type characters by pressing the corresponding keys or select the new comment text from the menu 2 Press Del to del...

Страница 199: ...he system will clear all comments on the image Press the user defined key for Delete Text or long press Clear key to delete all the comments 13 2 5 Setting the Start Position Perform the following pro...

Страница 200: ...between the separate image windows Perform the following procedure 1 Press Cursor and use the trackball trackpad to move the cursor onto the body mark The cursor then becomes indicating you can move...

Страница 201: ...to external memory devices For a save image you can perform operations like image reviewing analyzing and demonstration 14 1 1 Storage Media System supported memory media including System hard disk US...

Страница 202: ...o the system quickly Press the user defined key for Save Cine Retrospective Prospective to save the cine file in the default file directory in the CIN format The thumbnail of this image will appear in...

Страница 203: ...d patient display at the bottom of the screen When you move the cursor onto a thumbnail its name and format will display On the Review page the thumbnails refer to the images stored in the same exam W...

Страница 204: ...tus 2 To exit the image analysis Press Freeze to exit and enter the real time scan status Press Return to exit from the image analysis to the Review status In image analysis status the selected image...

Страница 205: ...selected the patient information and images are removed If Remove Images is selected only the patient images are removed 4 Select whether to hide the patient information 5 Select whether to encrypt ba...

Страница 206: ...rt Report on the screen 2 Select report type to be exported 3 Click OK to confirm The size of the report can be set see 4 9 Print Preset 14 3 Patient Data Management iStation The patient data include...

Страница 207: ...m of the screen 14 3 2 Patient Data View Management Select the desired patient information in the list Item Description Review an image Select an exam of a patient click Review Image to enter Review s...

Страница 208: ...xam Click to import the patient data from an external media Send Exam You can use this function to export the exam data to external devices in PC data or DICOMDIR data format and then import to PC or...

Страница 209: ...enabled Only the whole number ranging from 1 to 365 can be input 2 Click Modify 14 5 iStorage TIP To use iStorage function you need UltraAssist software in 2 0 version with V1 0 network protocol consu...

Страница 210: ...ports the PC format transfer function while CD DVD writing using Back Up supports only system relevant formats The symbol indicates that the input CD DVD is damaged or contains data in an incorrect fo...

Страница 211: ...r images in the iStation or Review screen Click Send Exam for the image iStorage task In iStation screen select the target exam and click Send Exam then click iStorage in the menu which appears MedTou...

Страница 212: ...system brings up a dialog box for entering the IP address of the remote server 2 Enter the IP address and click OK 3 Log on with the server account and password 4 Check the data transferred and carry...

Страница 213: ...rage Commitment and Query Retrieve applications For detailed information about DICOM presets see 4 7 DICOM HL7 15 1 DICOM Storage DICOM Storage is used to send images single frame or multi frame or st...

Страница 214: ...ee 4 1 7 Key Configuration Start the ultrasound exam scan Press the user defined key to send the image or the cine to DICOM storage 15 1 3 To send images to storage after an exam ends TIP To preset Se...

Страница 215: ...t of the screen and then click to select a thumbnail Or select an exam or exams from the patient list there should be images for this exam Tap Review on the exam main screen to enter the Review screen...

Страница 216: ...e keyword type enter the keywords and then click Query to search To reset the criteria click Clear button 6 Select the desired patient from the list Click Start Exam the patient information is importe...

Страница 217: ...he storage server Meanwhile it will send storage commitment to the storage commitment server 15 5 2 To send storage commitment automatically after an exam ends TIP Preset Sending printing after End Ex...

Страница 218: ...DVD optical file should be UDF and the optical disk should not be damaged Normally read write data from the USB ports on the ultrasound system File system format of removable device USB flash drive is...

Страница 219: ...ocedures DICOM structured report installed with the corresponding exam mode The state of the SR should be Cancel or Stop Set the storage option to Attach SR When Store Images or Only Store SR see 4 7...

Страница 220: ...This page intentionally left blank...

Страница 221: ...cology Obstetrics Cardiac Vascular Thoracic Pleural Urology Small Parts C5 1s Body surface Abdomen Gynecology Obstetrics Vascular Nerve Musculo skeletal Urology Thoracic Pleural Small Parts Supports c...

Страница 222: ...V11 3s Intracavitary Gynecology Obstetrics Urology Supports contrast imaging under Gynecology and Urology exam modes Supports strain elastography under Gynecology exam mode V11 3Hs Intracavitary Gyne...

Страница 223: ...function L12 3RCs Body surface Abdomen Pediatric Small Parts Musculo skeletal Vascular Thoracic Pleural Supports contrast imaging under Small Parts exam mode Supports strain elastography under Small P...

Страница 224: ...exam mode L20 5s Body surface Abdomen Small Parts Musculo skeletal Vascular Nerve Ophthalmic Supports contrast imaging under Small Parts exam mode Supports strain elastography under Small Parts and Mu...

Страница 225: ...ode P8 2s Body surface Abdomen Pediatric Cardiac Cephalic Supports TDI Stress Echo and Tissue Tracking under Cardiac exam mode P10 4s Body surface Abdomen Pediatric Cardiac Nerve Cephalic Supports TDI...

Страница 226: ...ng the sound beams in a given direction meanwhile it receives the reflected ultrasonic signal and converts it into an electrical signal for transmission over the cable The lens on the surface is the a...

Страница 227: ...guided biopsy procedure is performed Failure to do so may cause the probe and the needle guided bracket becomes a source of infection This section describes general procedures for operating the probe...

Страница 228: ...em Examinations Wiping off the ultrasound gel Thoroughly cleaning the probe Drying the probe Disinfecting the probe Rinsing the probe Drying the probe Inspection after use Storage Biopsy procedure Ste...

Страница 229: ...not be sufficient DO NOT use such a probe sheath The cover contains natural rubber latex and talc that can cause allergic reactions in some individuals DO NOT use an expired probe sheath Before using...

Страница 230: ...First Street South Kalona IA 52247 9589 USA Tel 1 319 656 4447 E mail info civco com http www civco com Perform the following procedure to install the probe sheath 1 Place an appropriate amount of gel...

Страница 231: ...use the probe to become deformed or damaged Observe the graph here carefully to immerse the probe Only soak parts of the probe below the strain relief Repeated disinfection will eventually damage the...

Страница 232: ...ur hospital policy and procedures for cleaning Perform the following procedure 1 Wear a pair of gloves to prevent infection 2 Disconnect the probe from the system If the sheath is used take off the sh...

Страница 233: ...r Observe the graph here carefully to perform disinfection Do not spray the strain relief on the connector end or the connector 4 Wipe away the residual disinfectant on the probe by using a piece of l...

Страница 234: ...e residual disinfectant Repeat the operation twice Or follow the disinfectant manufacturer s instructions regarding rinsing 5 Dry the probe by wiping with a piece of clean disposable lint free soft cl...

Страница 235: ...soft cloth Do not dry the probe by heating 4 Check whether the probe has defects such as peeling rifts bumps cracks or liquid spill If such defects exist the probe has reached the end of its service...

Страница 236: ...erating 0 C 40 C 20 85 700 hPa 1060 hPa Storage and transportation 20 C 55 C 20 95 700 hPa 1060 hPa C11 3s Operating 0 C 40 C 30 85 700 hPa 1060 hPa Storage and transportation 20 C 55 C 30 95 700 hPa...

Страница 237: ...a L20 5s Operating 0 C 35 C 15 80 700 hPa 1060 hPa Storage and transportation 20 C 60 C 15 90 500 hPa 1060 hPa 7LT4s Operating 0 C 40 C 25 90 550 hPa 1060 hPa Storage and transportation 20 C 60 C 10 9...

Страница 238: ...biopsy needle may fail to penetrate the target The incorrect biopsy may cause various side effects in the patient Use a needle guided bracket other than that provided Mount the needle guided bracket...

Страница 239: ...displayed on the ultrasound image does not indicate the actual position of the biopsy needle Therefore it should only be used as a reference Always monitor the relative positions of the biopsy needle...

Страница 240: ...ed Brackets Available Some of the probes have matched needle guided brackets for biopsy the available probes and the corresponding needle guided brackets are listed as follows Probe Needle Ultrasound...

Страница 241: ...16G 17G 18G L9 3s NGB 034 Metal needle detachable 40 50 60 14G 16G 18G 20G 22G C4 1s NGB 036 Metal needle detachable 7 25 35 14G 16G 18G 20G 22G L12 3RCs NGB 043 Metal needle detachable Angle range 3...

Страница 242: ...needle guided bracket immediately stop using it and contact MINDRAY Customer Service Department or sales representative Sterilize the needle guided bracket before and after use Put on the sterile prob...

Страница 243: ...en turn the retaining clamp to align it with the probe c When the retaining clamp is turned to the correct position the locking nut will lock the retaining clamp and the needle guided bracket is then...

Страница 244: ...hand Match the groove and tab with the tab and groove of the probe respectively Mount the bracket onto the probe c Screw the pinch nut of the needle guided bracket to ensure that the needle guided bra...

Страница 245: ...k and push it into the groove above the angle block then clamp it tightly b Screw the block s nut to secure the block c Insert a biopsy needle with the same specification as that of the guiding block...

Страница 246: ...terile probe sheath b Hold the probe by one hand select the proper needle guided bracket and hold it with the other hand Align the narrow end tab of the needle guided bracket with the groove of the pr...

Страница 247: ...e guiding block a Select a proper guiding block and push it into the groove above the angle block and clamp it tightly b Insert a biopsy needle with the same specification as that of the guiding block...

Страница 248: ...NGB 010 Metal needle detachable needle guided bracket Perform the following procedure 1 Install the needle guided bracket a Put on the sterile probe sheath 1 Guiding block 2 Guiding hole of the biopsy...

Страница 249: ...le guided bracket to match with the tabs of the probe Set the needle guided bracket at the desired position turn tightly the knob of fixing needle guided bracket to fix the needle guided bracket d Che...

Страница 250: ...edle guided bracket a Put on the sterile probe sheath b Connect the locating groove on the clamp with the two raised edges on the probe head and align the locating pit of the clamp with the convex poi...

Страница 251: ...ount the bracket onto the probe c Screw the pinch nut of the needle guided bracket to confirm that the needle guided bracket is properly installed on the probe 2 Adjust the needle angle to the proper...

Страница 252: ...ck pin and close the V shaped cover to fix the lock pin in the groove of the needle type adjusting base so as to install the needle into the guiding hole 4 Release the needle from the bracket a Pull t...

Страница 253: ...onto the probe c Screw the pinch nut of the needle guided bracket to ensure that the needle guided bracket is properly installed on the probe 2 Adjust the needle angle to the proper shift as required...

Страница 254: ...o the guiding block hole 4 Release the needle from the bracket a Loosen the guiding block s nut and slightly move the guiding block in the direction of the needle s tail b Separate the residual part o...

Страница 255: ...taining nut to fix the bracket and the probe 2 Adjust the needle angle to the proper shift as required a Loosen the angle pinch nut b Adjust the angle block to the desired level c Tighten the angle pi...

Страница 256: ...nd the needle leans to V shaped block c Hold the probe and release the pressure position of the needle Adjust the needle type adjusting nut manually following the direction of the arrow The needle mov...

Страница 257: ...from the needle 5 Remove the needle guided bracket a Rotate the clamping nuts of the needle guided bracket on right and left side following the direction of the arrow The needle guided bracket is sep...

Страница 258: ...groove c Push the biopsy forward arrow s direction until the locating pole inserting into the location hole Turn the lower clamp against the intra cavity probe Tighten the nut to lock the biopsy arro...

Страница 259: ...Metal needle detachable needle guided bracket Perform the following procedure 1 Install the needle guided bracket a Put on the sterile probe sheath 1 Needle fixing nut 2 Guiding groove 3 Angle fixing...

Страница 260: ...ired level c Tighten the angle pinch nut 3 Insert the biopsy needle a Adjust the dial scale to the required needle type shift b Pull the lock pin and close the V shaped cover to fix the lock pin in th...

Страница 261: ...s of the probe c Mount the bracket onto the probe and rotate the locking nut to fix the bracket 2 Insert the biopsy needle a Insert a biopsy needle into the guiding block hole b Adjust the biopsy angl...

Страница 262: ...the probe b Hold the probe and take out the bracket NGB 044 Metal needle detachable needle guided bracket Perform the following procedure 1 Install the needle guided bracket 1 Right clamp 2 Groove 3...

Страница 263: ...abs of the probe c Mount the bracket onto the probe and rotate the locking nut to fix the bracket 2 Adjust the needle angle to the proper shift as required a Loosen the depth pinch nut b Adjust the de...

Страница 264: ...to the guiding block hole 4 Remove the needle guided bracket a Hold the probe and remove the needle b Loosen the guiding block nut and remove it from the guiding block fixing frame c Rotate the clampi...

Страница 265: ...the needle guided bracket a Put on the sterile probe sheath b Align the needle guided bracket with the probe to align the locating bulges on the needle guide with the locating grooves on the probe th...

Страница 266: ...he probe and insert the needle into Needle entry of the needle guide 2 Remove the needle guided bracket a Hold the probe and remove the needle b Hold the probe and raise the needle guided bracket to s...

Страница 267: ...gle pinch nut 3 Insert the biopsy needle a Hold the probe Press the separation pressure position to separate needle guided V shaped block from pressure position of the needle b Put the needle into the...

Страница 268: ...sy to enter Biopsy Select the biopsy bracket angle guide line If the needle guided bracket supports more than one biopsy angle select the angle guideline by using Biopsy Kit Select the guide line dot...

Страница 269: ...ctual situation Tap Verify to enter Biopsy Verify menu to fine tune the guide line if needed 4 If available use iNeedle function to help enhance the needle visualization when the needle display is not...

Страница 270: ...e g liquinox MetriZyme by using distilled or softened water in accordance with the operator s manual provided by the manufacturer 4 Detach all the detachable parts of the needle guided bracket and imm...

Страница 271: ...guided bracket inside a high temperature steam sterilizer and perform sterilization The sterilization parameters are 121 C and 30 minutes for a gravity displacement steam sterilizer and are 132 C and...

Страница 272: ...trackball trackpad to change the mark position and by adjusting lithotripsy machine tools or patient posture to locate the stone center at the mark b Read the depth of the mark by observing the depth...

Страница 273: ...professional the eSpacial Navi system can achieve precision as high as 2 4 mm Place the magnetizer in the proper position to prevent the magnetizer from falling off or becoming damaged DO NOT make the...

Страница 274: ...ype 5 Scale of guideline 5 mm scale 6 Projection position of the needle guidance trajectory in the ultrasound plane 7 Position at which the needle cross the ultrasound plane 8 Needle projection positi...

Страница 275: ...eS Navi Needle menu Do not move the probe during the calibration 3 Magnetize a needle cap a Place an appropriate sterile needle cap length 3 5cm to 4 5cm diameter 7 mm with bottom closed into the mag...

Страница 276: ...the maximum value the needle guidance information disappears temporally 2 Adjust the needle position and angle to define the position for inserting the needle Pay attention to the field strength durin...

Страница 277: ...with a disinfectant Follow the disinfectant manufacturer s recommended contact time and mode Table 16 7 Compatible disinfectants Trade Name Manufacturer Chemical Name Procedures Surfa safe Anios Labor...

Страница 278: ...of water about 2 gallons for at least 1 minute to remove all chemical residues on it Or follow the rinsing method recommended by the disinfectant manufacturer to rinse the magnetizer 4 Wipe off the wa...

Страница 279: ...save the recording file automatically you can select to save in local disk U disk or optical disk Perform the following procedure 1 Perform ultrasound exams select appropriate views and adjust paramet...

Страница 280: ...nding progress the icon displays as 17 3 DVR Video Replay You can replay the video and audio record 17 3 1 Replay on PC Connect the USB disk or optical disk with the file to the PC open the file direc...

Страница 281: ...aintenance 18 1 1 Cleaning the System WARNING For the system equipped with trackball before cleaning the system be sure to turn off the power and disconnect the power cord from the outlet If you clean...

Страница 282: ...to air dry Perform the following procedure 1 Use a soft dry cloth to wipe away dust attached to the inside outside and gaps in the probe holder Use a soft brush to brush away dust or stains from the...

Страница 283: ...mall ball otherwise the small ball may drop down When liquid is spray onto the trackball area most of the liquid can be discharged through the waterspout besides you can dry the leaving water using th...

Страница 284: ...controlling infection For the method to control infection consult your hospital s infection control officer or epidemiologist CAUTION Never immerse any part of the main unit in liquids or allow liquid...

Страница 285: ...elect to lock the system for 10 seconds to disinfect the system while the power is On After cleaning disinfect the system as follows 1 Wear medical gloves to prevent infection 2 Disinfect the main uni...

Страница 286: ...he connector pins are bent destroyed or falling off Content Description Color and B W video printer First wipe off dust or stain attached to the cover of printer with soft dry cloth then clean the ins...

Страница 287: ...e power indicator of the monitor is lit up but the image is blank The interval between turning off and restarting the system is too short wait at least 20 seconds Turn off the system and wait at least...

Страница 288: ...This page intentionally left blank...

Страница 289: ...low power visible LED DO NOT stare into beam because of unknown hazards of transient radiation provided by class 2 laser DO NOT stare into beam emitted by SYMBOL DS4308 for more than 10 s CAUTION Ens...

Страница 290: ...ector is properly secured 2 Connect the other end of the interface cable to the host A 1 2 Setting The reader has factory settings please refer to A 4 for details The reader supports some user defined...

Страница 291: ...ameter Low Volume Medium Volume High Volume Code 93 and codebar scanning To enable or disable Code 93 scan the appropriate barcode below To enable Codebar scan the appropriate barcode below Code 39 fu...

Страница 292: ...decode only I 2 of 5 symbols with 8 characters scan I 2 of 5 One Discrete Length then scan 0 followed by 8 A 1 3 Scanning in Hand Held Mode Perform the following procedure 1 Ensure all connections ar...

Страница 293: ...y back into the reader from the barcode is known as specular reflection This specular reflection can make decoding difficult You can tilt the reader up to 55 forward or back and achieve a successful d...

Страница 294: ...t provided Perform Scanning in Hands Free Mode When the reader is seated in the scanner cup the reader s built in sensor places the reader in hands free mode When you remove the reader from the stand...

Страница 295: ...ple The 2 D barcode reader supports hand held operation mode Hand held mode press the trigger to decode A 2 1 Overview 1 LED Green A barcode was successfully decoded Red A data transmission error or r...

Страница 296: ...t the other end of the interface cable to the host Removing the Interface Cable Perform the following procedure 1 Using the tip of a screwdriver or some other tools with a sharp head depress the cable...

Страница 297: ...me Code 93 and codabar setting To enable Code 93 scan the appropriate barcode below To enable Codabar scan the appropriate barcode below Code 39 full ASCII setting Code 39 Full ASCII is a variant of C...

Страница 298: ...ct the length using the numeric barcodes below For example to decode only I 2 of 5 symbols with 8 characters scan I 2 of 5 One Discrete Length then scan 0 followed by 8 A 2 4 Scanning in Hand Held Mod...

Страница 299: ...gital imager reader beeps indicating the barcode is successfully decoded Steps 2 4 above may be required to repeat on poor quality or difficult barcodes The aiming pattern is smaller when the digital...

Страница 300: ...ropriate interface cable to the barcode reader 3 Plug the other end of the cable into a free USB port on the Ultrasound system 4 Once the imager has been fully connected power on the Ultrasound System...

Страница 301: ...is properly connected to the Ultrasound System before scanning After the barcode is successfully scanned the barcode reader buzzes and the green indicator is On Configuration before use Perform the f...

Страница 302: ...Wipe the window using a tissue moistened with ammonia water Do not spray water or other cleaning liquids directly into the window A 5 Parameter Defaults Refer to the following table for parameter def...

Страница 303: ...Disable UPC E1 Preamble System Character Code 128 Emulation Disable Convert UPC E to A Disable Data Matrix Enable Convert UPC E1 to A Disable Maxicode Enable EAN 8 JAN 8 Extend Disable QR Code Enable...

Страница 304: ...This page intentionally left blank...

Страница 305: ...result Be sure to connect the equipotential wire before inserting the power plug into the receptacle be sure to remove the power plug from the receptacle before disconnecting the equipotential wire o...

Страница 306: ...t AC supply 7 Power outlet Supplies the power for peripheral devices 8 Power adapter bracket Used for placing the power adapter 9 Lifting column Used for adjusting the height of operation panel 10 Pro...

Страница 307: ...nk purchased separately instead NOTE Keep the main unit tray carefully for future use after it is removed from the trolley Reinstall the main unit tray to the trolley if the U Bank has been removed fr...

Страница 308: ...This page intentionally left blank...

Страница 309: ...An electrical safety inspection should be periodically performed every two years The safety analyzer is also an excellent troubleshooting tool for detecting abnormalities in line voltage and groundin...

Страница 310: ...Earth Leakage Test Run an Earth Leakage test on the device being tested before performing any other leakage tests The following outlet conditions apply when performing the Earth Leakage test normal p...

Страница 311: ...ngle Fault Condition reverse polarity with open earth Single Fault Condition LIMITS For ES 60601 1 100 A in Normal Condition 300 A in Single Fault Condition For IEC 60601 1 100 A in Normal Condition 5...

Страница 312: ...Polarity Reversed Polarity LIMITS For BF applied parts 5000 A C 9 Patient Auxiliary Current Patient Auxiliary currents are measured between any selected Applied Part connector and the remaining Applie...

Страница 313: ...d excess the router capacity please refer to settings of the router generally it should be 5 DO NOT connect devices other than specified into the LAN Medical devices within the same LAN may interfere...

Страница 314: ...or after successful connection 5 Click Refresh to refresh the Wireless Network Connection list D 2 IP Configure NOTE When the system background is processing network task DICOM sending for example ple...

Страница 315: ...er If Static is selected using a static IP address enter the IP address IP address of the system should be in the same network segment with the server 2 Click OK to save current setting Click Cancel t...

Страница 316: ...This page intentionally left blank...

Страница 317: ...lable under abdominal gynecological urological obstetrical small parts and nerve block area E 1 Use iScanHelper for Reference Perform the following procedure 1 Perform ordinary scanning procedure 2 Pr...

Страница 318: ...or Ultrasonic image to go to the single window of each of them Double click single window again to return to quad window display 1 Section selecting box Move the cursor onto the target view and press...

Страница 319: ...mode 3 Select an item in the list and click Start to begin the demonstration 4 Click Exit to exit iVision status F 1 Demonstration item Demonstration items are image files in formats supported by the...

Страница 320: ...delete selected files or catalogs from the file list Clear to clear all the files or catalogs in the file list Export to export selected directories files to external storage devices Click Export to b...

Страница 321: ...n off Power Mode M Mode Turn on off M Mode PW Mode Turn on off PW Mode 3D Turn on off 3D Mode Freeze Freeze Image UnFreeze UnFreeze Image Clear Clear Dual Enter Dual Windows Format Quad Enter Quad Win...

Страница 322: ...Switch Exam Mode to Abdomen Vascular Neonatal Abdomen Switch Exam Mode to Neonatal Abdomen Adult Cardiac Switch Exam Mode to Adult Cardiac Cardiac Difficult Switch Exam Mode to Adult Cardiac Difficult...

Страница 323: ...to Emergency Obstetrics Emergency Vascular Switch Exam Mode to Emergency Vascular Emergency Superficial Switch Exam Mode to Emergency Superficial Lung Switch Exam Mode to Lung Neonatal Head Switch Exa...

Страница 324: ...of Vocal Commands Close Report Dialog Close Report Dialog Open Preset Dialog Open Preset Dialog Close Preset Dialog Close Preset Dialog Smart IVC Smart IVC Smart VTI Smart VTI Smart B lines Smart B l...

Страница 325: ...nostic ultrasound instruments the possibility exists that such biological effects may be identified in the future Thus ultrasound should be used in a prudent manner to provide medical benefit to the p...

Страница 326: ...s MI is an index that shows the possibility of the cavitations generation based on acoustic pressure and the value in which the peak rarefactional acoustic pressure is divided by the square root of th...

Страница 327: ...coustic power percentage the greater the current acoustic output When the image is frozen the system stops transmitting acoustic power Default setting of acoustic power Selection of diagnostic applica...

Страница 328: ...overall depth and PRF The operating mode determines whether the ultrasound beam is scanning or non scanning Thermal bioeffect is closely connected to M mode Doppler and Color mode Acoustic attenuation...

Страница 329: ...setting of the system increases or decreases the possibility of Thermal or Mechanical effect More specifically they were designed to assist in the implementation of the ALARA principle As an operator...

Страница 330: ...ipment Part 2 37 Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment issued by IEC in 2015 IEC 62359 Ultrasonics Field char...

Страница 331: ...should be avoided because it could result in improper operation If such use is necessary this equipment and the other equipment should be observed to verify that they are operating normally Use of ac...

Страница 332: ...in such an environment EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIROMENT GUIDANCE RF emissions CISPR 11 Group 1 The system uses RF energy only for its internal function Therefore its RF emissions...

Страница 333: ...E AND MINDRAY DECLARATION ELECTROMAGNETIC IMMUNITY The system is intended for use in the electromagnetic environment specified below The customer or the user of system should assure that it is used in...

Страница 334: ...ains interruptions it is recommended that our product be powered from an uninterruptible power supply or a battery Power frequency 50 60 HZ magnetic field IEC 61000 4 8 30 A m 30 A m Power frequency m...

Страница 335: ...ould be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 10 V m 80MHz 2 7GHz 10 V m...

Страница 336: ...ers at a maximum output power not listed above the recommended separation distanced in meters m can be determined using the equation applicable to the frequency of the transmitter where P is the maxim...

Страница 337: ......

Страница 338: ...P N 046 018835 00 2 0...