1 Safety Precautions
Operator’s Manual
1 - 7
•
It is required to practice ALARA when operating ultrasound system.
Minimize the acoustic power without compromising the quality of images.
•
The probe and accessories supplied with it are not delivered disinfected or
sterilized. Sterilization (or high-level disinfect) before use is required.
•
Disposable components should be packaged sterile and for single-use
only. Do not use if integrity of packaging violated or if expiration date has
passed. Please use the disposable components compliant with the relevant
local regulations.
•
Please use the disinfection or sterilization solution recommended in this
operator's manual; otherwise Mindray will not be liable for damage caused
by other solutions. If you have any questions, please contact Mindray
Customer Service Department or sales representative.
•
Do not use pre-lubricated condoms as a sheath. Lubricant may not be
compatible with the probe material and damage may result.
•
The damage of the transducer may be caused by the contact of improper
gel or cleaner:
– DO NOT dip the transducer in the strong polar solution of ethanol,
chloride of lime, ammonium chloride, acetone and formaldehyde.
– DO NOT contact the transducer with solution or ultrasound gel
containing oily medium such as mineral oil or lanoline.
NOTE:
•
Read the following precautions to prevent the probe from malfunction:
–
Before connecting or disconnecting the probe, freeze or turn off the diagnostic ultrasound
system.
–
Clean and disinfect the probe before and after each examination.
–
After the examination, wipe off the ultrasound gel thoroughly. Otherwise, the ultrasound
gel may solidify and the image quality would be degraded.
•
Repeated disinfection will eventually damage the probe, please check the probe performance
periodically.
1.3
Latex Alert
When choosing a probe sheath, it is recommended that you directly contact CIVCO for obtaining
information regarding probe sheaths, pricing, samples and local distribution.
For CIVCO information, please contact the following:
CIVCO Medical Instruments
Tel: 1-800-445-6741
www.civco.com
WARNING
Allergic reactions in patients sensitive to latex (natural rubber) may range from
mild skin reactions (irritation) to fatal anaphylactic shock, and may include
Содержание Ana
Страница 2: ......
Страница 50: ...This page intentionally left blank...
Страница 60: ...This page intentionally left blank...
Страница 110: ...This page intentionally left blank...
Страница 116: ...This page intentionally left blank...
Страница 166: ...This page intentionally left blank...
Страница 176: ...This page intentionally left blank...
Страница 194: ...This page intentionally left blank...
Страница 220: ...This page intentionally left blank...
Страница 288: ...This page intentionally left blank...
Страница 304: ...This page intentionally left blank...
Страница 308: ...This page intentionally left blank...
Страница 316: ...This page intentionally left blank...
Страница 337: ......
Страница 338: ...P N 046 018835 00 2 0...