16 Probes and Biopsy
Operator’s Manual
16 - 25
2.
Adjust the needle angle to the proper shift as required:
a.
Loosen the angle pinch nut.
b.
Adjust the angle block to the desired level.
c.
Tighten the angle pinch nut.
3.
Install the guiding block
a.
Select a suitable guiding block and push it into the groove above the angle block, then
clamp it tightly
b.
Screw the block's nut to secure the block.
c.
Insert a biopsy needle with the same specification as that of the guiding block into the
guiding block hole.
4.
Release the needle from the bracket:
a.
Screw the nut of the guiding block and remove the guiding block slightly along the
direction of the needle’s tail.
b.
Separate the residual part of the needle-guide bracket and the probe from the needle.
Содержание Ana
Страница 2: ......
Страница 50: ...This page intentionally left blank...
Страница 60: ...This page intentionally left blank...
Страница 110: ...This page intentionally left blank...
Страница 116: ...This page intentionally left blank...
Страница 166: ...This page intentionally left blank...
Страница 176: ...This page intentionally left blank...
Страница 194: ...This page intentionally left blank...
Страница 220: ...This page intentionally left blank...
Страница 288: ...This page intentionally left blank...
Страница 304: ...This page intentionally left blank...
Страница 308: ...This page intentionally left blank...
Страница 316: ...This page intentionally left blank...
Страница 337: ......
Страница 338: ...P N 046 018835 00 2 0...