16 Probes and Biopsy
Operator’s Manual
16 - 27
c.
Check manually to confirm that the needle-guided bracket is securely installed on the
probe.
2.
Install the guiding block:
a.
Select a proper guiding block and push it into the groove above the angle block, and
clamp it tightly.
b.
Insert a biopsy needle with the same specification as that of the guiding block into the
hole of the guiding block.
3.
Release the needle from the bracket:
a.
Remove the guiding block slightly along the direction of the needle’s tail.