Operator’s Manual
10 - 1
10
Physiological Unit Signal
The physiological unit signal waveform is used for checking ultrasound image in ultrasound exam
(cardiac exam mainly).
WARNING
•
Do not use the physiological traces for diagnosis and monitoring.
•
To avoid electric shock, the following checks shall be performed prior to an
operation:
– The ECG electrode cable must not be cracked, frayed or show any
signs of damage or strain.
– The ECG electrode cable must be correctly connected.
– You must use the ECG leads provided with the physiological unit
module. Failure to do so may result in electric shock.
•
The ECG electrode cable must be connected to the system first. Only after
the cable is connected to the system, can the patient be connected to the
ECG electrodes. Failure to follow these instructions may subject the patient
to electric shock.
•
Do not place the ECG electrodes directly in contact the patient’s heart;
otherwise it may lead to stop of the patient’s heartbeat.
•
Do not apply the ECG electrodes if the voltage exceeds 15 volts. This could
produce an electric shock.
•
Do not use this system when any digital device such as a high-frequency
electrotome, high-frequency therapeutic device or defibrillator is applied
already.
•
Conductive parts of electrodes and associated connectors for ECG should
not contact other conductive parts including earth/grounding.
•
Frequent trampling or squeezing on the cables may result in cable break-
down or fracture.
•
Display effect of respiratory curve depends on the patient breathing status.
While a very slow or smooth breathing may lead to an inapparent
respiratory curve, breathing in a large amplitude may cause an incomplete
display of the respiratory curve. Display effect is linked to the connected
parts of the body. Generally, signals by connecting to limbs are stronger
than by connecting to the chest.
10.1 ECG
Perform the following procedure:
Содержание Ana
Страница 2: ......
Страница 50: ...This page intentionally left blank...
Страница 60: ...This page intentionally left blank...
Страница 110: ...This page intentionally left blank...
Страница 116: ...This page intentionally left blank...
Страница 166: ...This page intentionally left blank...
Страница 176: ...This page intentionally left blank...
Страница 194: ...This page intentionally left blank...
Страница 220: ...This page intentionally left blank...
Страница 288: ...This page intentionally left blank...
Страница 304: ...This page intentionally left blank...
Страница 308: ...This page intentionally left blank...
Страница 316: ...This page intentionally left blank...
Страница 337: ......
Страница 338: ...P N 046 018835 00 2 0...