11 - 4
Operator’s Manual
11 Stress Echo
11.3 Review/WMS Mode
Review/WMS mode is used by cardiologists to evaluate clips for cardiac wall motion
abnormalities. Different views from different stages are selected for comparison across a wide
variety of combinations. The most common workflow is to compare “same views” but at “different
stages” of the exam (e.g., PSLA view, Rest stage compared to PSLA view, Post-exercise stage).
11.3.1 Enter review mode
Select [Review/WMS] to enter review mode, and then select the label of the phase or view (for
example, Rest or Long Axis), the system displays all loops that represent the selected phase or
view.
To display phases for the selected view(s)
Perform the following procedure:
1.
To include a phase or view for display, select the leftmost, gray box to the left of each required
phase and/or view.
The system inserts a checkmark into each selected gray box.
2.
To exclude a phase or view from display, select the blue box to the left of each required phase
and/or view.
The system inserts a X into each selected box, like
.
3.
Select [Display Selected] on the touch screen.
The system displays the selected phases for each selected view side by side.
To display all views for a specific phase
Select the phase label (for example, Rest).
To display all phases for a specific view
Select the view label (for example, PLAX).
Text
Function that turns the screen graphic text “On” or “Off”. Information
includes: name of level, name of view, heart rate, time stamp
acquisition, timers, frame slider, loop ID, clip control. For the cine
without distributed view, the name of level and name of view are
displayed in “--”.
Apply edit all
Clip edit applied to all clips taken.
Clip Length
Specify the clip segments: systole, diastole, full cycle or user-defined.
Bookmark
For continuous acquisition, when the bookmark is set to “On”, only the
selected loops for the current view can be displayed.
Delete Unselected
Delete clips that are not selected. If selected, the system will delete all
clips that are not selected after the exam is ended.
Suspend exam
Pauses the stress echo exam but does not end the stress echo exam.
When a stress echo exam is suspended, the user can perform image
acquisition of all other imaging modes, or perform operations such as
measurement.
End SE exam
End the stress echo exam.
Selection
Description
Содержание Ana
Страница 2: ......
Страница 50: ...This page intentionally left blank...
Страница 60: ...This page intentionally left blank...
Страница 110: ...This page intentionally left blank...
Страница 116: ...This page intentionally left blank...
Страница 166: ...This page intentionally left blank...
Страница 176: ...This page intentionally left blank...
Страница 194: ...This page intentionally left blank...
Страница 220: ...This page intentionally left blank...
Страница 288: ...This page intentionally left blank...
Страница 304: ...This page intentionally left blank...
Страница 308: ...This page intentionally left blank...
Страница 316: ...This page intentionally left blank...
Страница 337: ......
Страница 338: ...P N 046 018835 00 2 0...