6 - 14
Operator’s Manual
6 Image Acquisition
8.
Double press <Set> to finish the editing, and the sample line is displayed in white.
9.
You can edit the curve if needed:
a.
After finishing the sampling line, tap [Edit] on the touch screen. The cursor becomes the
icon .
b.
Move the cursor over the curve, press <Set> to activate the spot.
c.
Move the cursor to change the shape of the curve.
d.
Double press <Set> to finish the editing, and the sample line is displayed in white.
NOTE:
Tap [Delete] on the touch screen to remove the curve, and tap [Edit] to re-edit the curve.
10. Adjust the parameters to obtain the desired tissue of Free Xros CM image. Then, save the
image.
11. Tap [Free Xros CM] or press the user-defined key for “Free Xros CM”, or press <B> to exit.
6.7.3 Anatomical M Mode Parameters
In anatomical M mode, adjustable parameters are similar with these in M mode.
6.8
PW/CW Mode
PW (Pulsed Wave Doppler) mode or CW (Continuous Wave Doppler) mode is used to provide
blood flow velocity and direction utilizing a real-time spectral display. The horizontal axis
represents time, while the vertical axis represents Doppler frequency shift.
PW mode provides a function to examine flow at one specific site for its velocity, direction and
features; while CW mode proves to be much more sensitive to high velocity flow display. Thus, a
combination of both modes will contribute to a much more accurate analysis.
CW module is optional.
6.8.1 PW/CW Mode Image Scanning
Perform the following procedure:
1.
Select a premium image during B mode or B+ Color (Power) scan, and adjust to place the area
of interest in the center of the image.
2.
Press <PW>/<CW> to adjust the sampling line.
The sampling status will be displayed in the image parameter area.
3.
Set the position of the sample line and the SVD by using the trackball/trackpad, and adjust the
angle and SV size according to the actual situation.
4.
Press <PW>/<CW> again or <Update> to enter PW/CW mode and perform the examination.
Observe and calculate the data based on B mode or Color mode image. You can also adjust the
SV size, angle and depth in the real-time scan.
5.
Adjust the image parameters to obtain optimized images.
6.8.2 PW/CW Mode Image Parameter
Gain
This function is intended to adjust the gain of spectrum map. The real-time gain value is displayed
in the image parameter area.
Содержание Ana
Страница 2: ......
Страница 50: ...This page intentionally left blank...
Страница 60: ...This page intentionally left blank...
Страница 110: ...This page intentionally left blank...
Страница 116: ...This page intentionally left blank...
Страница 166: ...This page intentionally left blank...
Страница 176: ...This page intentionally left blank...
Страница 194: ...This page intentionally left blank...
Страница 220: ...This page intentionally left blank...
Страница 288: ...This page intentionally left blank...
Страница 304: ...This page intentionally left blank...
Страница 308: ...This page intentionally left blank...
Страница 316: ...This page intentionally left blank...
Страница 337: ......
Страница 338: ...P N 046 018835 00 2 0...