16 - 26
Operator’s Manual
16 Probes and Biopsy
5.
Remove the needle-guided bracket:
a.
Unscrew the Needle-guided bracket pinch nut, and remove the needle-guided bracket
from the probe.
b.
Separate the probe and the needle-guided bracket.
NGB-007 Plastic/needle detachable needle-guided bracket
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Hold the probe by one hand, select the proper needle-guided bracket, and hold it with the
other hand. Align the narrow end tab of the needle-guided bracket with the groove of the
probe, then push the needle-guided bracket forward, making the tabs and the grooves of
the needle-guided bracket to match with the grooves and tabs of the probe.
1
Support of needle-guided bracket
2
Angle block
3
Guiding block
4
Groove and tab of the needle-guided bracket
5
Specification of guiding block
6
Guiding hole of the biopsy needle
7
Specification of angle block
Содержание Ana
Страница 2: ......
Страница 50: ...This page intentionally left blank...
Страница 60: ...This page intentionally left blank...
Страница 110: ...This page intentionally left blank...
Страница 116: ...This page intentionally left blank...
Страница 166: ...This page intentionally left blank...
Страница 176: ...This page intentionally left blank...
Страница 194: ...This page intentionally left blank...
Страница 220: ...This page intentionally left blank...
Страница 288: ...This page intentionally left blank...
Страница 304: ...This page intentionally left blank...
Страница 308: ...This page intentionally left blank...
Страница 316: ...This page intentionally left blank...
Страница 337: ......
Страница 338: ...P N 046 018835 00 2 0...