1. Insert the isokinetic probe’s barb connector into one end of the
tubing. If the tripod will be used, screw the probe onto the tripod’s
head and use the lock wheel to lock the probe in position. The head
of the tripod swivels and can be locked in any position. See Figure
3-10.
Figure 3-10 Tubing, isokinetic probe and tripod
2. Remove the protective cap from the sample inlet located on top of
the unit.
3. Install the nut, rear ferrule and front ferrule onto the tubing, if
needed. Insert the tubing into the inlet and make sure the two
ferrules contact the inlet as shown in Figure 3-11. Finger tighten the
nut carefully to prevent cross-threading. Do NOT overtighten or the
nut may crack.
Figure 3-11 Installing Tubing onto Inlet (Example)
WARNING:
Instrument ferrule style inlet caps must be the
Lighthouse supplied item, which is constructed of soft
nylon material. The use of metal stainless steel inlet caps is
discouraged as it can cause abrasion damage to the inlet
nozzle and particles can be drawn into the sensor. See
Figure 3-12.
Содержание SOLAIR 1100LD
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions SOLAIR 1100LD Airborne Particle Counter Gen E Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse SOLAIR 1100LD Operating Manual t vi 248083395 1 Rev 9...
Страница 14: ...Lighthouse SOLAIR 1100LD Operating Manual ii 248083395 1 Rev 9...
Страница 20: ...Lighthouse SOLAIR 1100LD Operating Manual 2 4 248083395 1 Rev 9...
Страница 58: ...Lighthouse SOLAIR 1100LD Operating Manual 4 24 248083395 1 Rev 9...
Страница 64: ...Lighthouse SOLAIR 1100LD Operating Manual 5 6 248083395 1 Rev 9...
Страница 146: ...Operate SOLAIR 248083395 1 Rev 9 6 82...
Страница 158: ...Lighthouse SOLAIR 1100LD Operating Manual A 2 248083395 1 Rev 9...
Страница 180: ...Lighthouse SOLAIR 1100LD Operating Manual D 2 248083395 1 Rev 9...
Страница 187: ...00...
Страница 188: ......