Lighthouse SOLAIR 1100LD Operating Manual
7-6
248083395-1 Rev 9
The low word register for Sample Time is 40034.
The low word register for Hold Time is 40032.
The low word register for Initial Delay is 40030.
Table 7-4 Instrument Parameters
Register
Data Type
Description
40026
unsigned integer
Location number; Specifies location of
Particle Counter.
40029
unsigned integer
Initial Delay [high]. Works in conjunction
with 40030. Number of seconds to wait
before starting the first sample. Max value
is 86,399, which equals 23h 59m 59s.
40030
unsigned integer
Initial Delay [low]
40031
unsigned integer
Hold Time [high]. Works in conjunction
with 40032. Number of seconds to wait
between sample periods. Max value is
86,399, which equals 23h 59m 59s
40032
unsigned integer
Hold Time [low]
40033
unsigned integer
Sample Time [high]. Works in
conjunction with 40034. Number of
seconds to sample. Max value is 86,399,
which equals 23h 59m 59s.
40034
unsigned integer
Sample Time [low]
Содержание SOLAIR 1100LD
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions SOLAIR 1100LD Airborne Particle Counter Gen E Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse SOLAIR 1100LD Operating Manual t vi 248083395 1 Rev 9...
Страница 14: ...Lighthouse SOLAIR 1100LD Operating Manual ii 248083395 1 Rev 9...
Страница 20: ...Lighthouse SOLAIR 1100LD Operating Manual 2 4 248083395 1 Rev 9...
Страница 58: ...Lighthouse SOLAIR 1100LD Operating Manual 4 24 248083395 1 Rev 9...
Страница 64: ...Lighthouse SOLAIR 1100LD Operating Manual 5 6 248083395 1 Rev 9...
Страница 146: ...Operate SOLAIR 248083395 1 Rev 9 6 82...
Страница 158: ...Lighthouse SOLAIR 1100LD Operating Manual A 2 248083395 1 Rev 9...
Страница 180: ...Lighthouse SOLAIR 1100LD Operating Manual D 2 248083395 1 Rev 9...
Страница 187: ...00...
Страница 188: ......