Lighthouse SOLAIR 1100LD Operating Manual
6-28
248083395-1 Rev 9
Alarm
The Particle Alarm screen allows the user to enable alarming on
specific channels as shown in Figure 6-37.
Figure 6-37 Particle Alarm Configuration Example
Note:
Alarming is only
applicable for AUTO and
MANUAL mode. It applies
only to Raw particle counts
even if the instrument is
displaying Normalized
data.
If the instrument is set to
Differential data, the alarm
threshold will apply to the
differential counts.
If the instrument is set to
display Cumulative data,
the alarm threshold will
apply to the cumulative
counts.
To enable the alarming for any channel, press the “
X
” next to that
channel. When a checkmark is displayed, that channel is enabled for
alarming. Press the checkmark to disable the alarming for that channel.
Alarm Threshold
Press the threshold button next to the enabled channel in order to set the
alarm threshold for that channel. The Alarm Threshold screen will open
as shown in Figure 6-38.
Figure 6-38 Configure Alarm Threshold
Threshold Button
Содержание SOLAIR 1100LD
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions SOLAIR 1100LD Airborne Particle Counter Gen E Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse SOLAIR 1100LD Operating Manual t vi 248083395 1 Rev 9...
Страница 14: ...Lighthouse SOLAIR 1100LD Operating Manual ii 248083395 1 Rev 9...
Страница 20: ...Lighthouse SOLAIR 1100LD Operating Manual 2 4 248083395 1 Rev 9...
Страница 58: ...Lighthouse SOLAIR 1100LD Operating Manual 4 24 248083395 1 Rev 9...
Страница 64: ...Lighthouse SOLAIR 1100LD Operating Manual 5 6 248083395 1 Rev 9...
Страница 146: ...Operate SOLAIR 248083395 1 Rev 9 6 82...
Страница 158: ...Lighthouse SOLAIR 1100LD Operating Manual A 2 248083395 1 Rev 9...
Страница 180: ...Lighthouse SOLAIR 1100LD Operating Manual D 2 248083395 1 Rev 9...
Страница 187: ...00...
Страница 188: ......