Lighthouse SOLAIR 1100LD Operating Manual
4-20
248083395-1 Rev 9
10.
Type
telnet xx.xx.xx.xx 9999
and press
Enter
. This command
accesses the telnet port of the instrument. The expected error
message from Step 9. is shown in Figure 4-18.
Figure 4-18 Command - Send “Telnet 9999"
The user will have 4 seconds to respond in the next step. If they
respond too slowly, they will have to repeat step 10.
11.
Press
Enter
to start the 1100LD Ethernet interface setup program.
Figure 4-19 displays some of the default settings - make no
changes to these settings except as instructed.
Figure 4-19 Starting the Set Up Program