Lighthouse SOLAIR 1100LD Operating Manual
6-74
248083395-1 Rev 9
Setting up Counter to Run Reports
For this example, the FED STD FT
3
report will be used. See
Figure 6-101.
Figure 6-101 FED STD FT3 Report Screen
•
Set the instrument for the Fed Std ft
3
report, 100 ft
2
, Class 1000,
unidirectional air flow, operational room status.
•
This will generate the following minimum requirements:
Note:
The Minimum
Volumes are measured in the
standard measurement
required by each specific
report.
The number of digits allowed
for area to calculate volume is
limited to a 5 digit value for ft2
and a 4 digit value for m2.
•
Min Locations: 4
•
Min Samples: 5
Table 6-1 Table - Minimum Volume per Channel Size
Ch Size (
P
)
Min Vol (ft
3
)
0.10
0.10
0.15
0.10
0.20
0.10
0.25
0.10
0.30
0.10
0.50
0.10
0.70
0.10
1.00
0.10
Содержание SOLAIR 1100LD
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions SOLAIR 1100LD Airborne Particle Counter Gen E Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse SOLAIR 1100LD Operating Manual t vi 248083395 1 Rev 9...
Страница 14: ...Lighthouse SOLAIR 1100LD Operating Manual ii 248083395 1 Rev 9...
Страница 20: ...Lighthouse SOLAIR 1100LD Operating Manual 2 4 248083395 1 Rev 9...
Страница 58: ...Lighthouse SOLAIR 1100LD Operating Manual 4 24 248083395 1 Rev 9...
Страница 64: ...Lighthouse SOLAIR 1100LD Operating Manual 5 6 248083395 1 Rev 9...
Страница 146: ...Operate SOLAIR 248083395 1 Rev 9 6 82...
Страница 158: ...Lighthouse SOLAIR 1100LD Operating Manual A 2 248083395 1 Rev 9...
Страница 180: ...Lighthouse SOLAIR 1100LD Operating Manual D 2 248083395 1 Rev 9...
Страница 187: ...00...
Страница 188: ......