Lighthouse SOLAIR 1100LD Operating Manual
4-6
248083395-1 Rev 9
Peripheral
Interface
Connector
The back of the SOLAIR also contains a round, Peripheral Interface
Connector (PIC) that can be used to connect an external peripheral
device such as an alarm buzzer, alarm light or a remote start/stop
controller.
Table 4-2 displays the pinouts for PIC connector.
Using Report
Parameters
The REPORTS screen is not just for printing reports. The data provided
on the screen when various parameters are set will allow users to setup
the counter in order to meet the requirements necessary for the report
used in the user’s Standard Operating Procedure.
Note:
Due to rounding of
values, occasionally the
user may need to increase
the sample volume in order
to meet the requirements.
For each report type, as the user changes the Room Area, the Class
Level, and sometimes the Air Flow or the Room Status, the values for
the Minimum Locations, Minimum Samples and Minimum Volume
may change and will be correctly calculated for the requirements for
each report type.
By confirming that the setup of the counter meets the requirements
indicated on this screen, the user will ensure that the samples run will
meet the requirements of the selected report.
Table 4-2 PIC Connector Pin-Outs
Pin
Connection
1
24 VDC
2
GND
3
Alarm+
4
Alarm-
5
Start/Stop
6
Not Connected
7
Not Connected
8
GND
Содержание SOLAIR 1100LD
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions SOLAIR 1100LD Airborne Particle Counter Gen E Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse SOLAIR 1100LD Operating Manual t vi 248083395 1 Rev 9...
Страница 14: ...Lighthouse SOLAIR 1100LD Operating Manual ii 248083395 1 Rev 9...
Страница 20: ...Lighthouse SOLAIR 1100LD Operating Manual 2 4 248083395 1 Rev 9...
Страница 58: ...Lighthouse SOLAIR 1100LD Operating Manual 4 24 248083395 1 Rev 9...
Страница 64: ...Lighthouse SOLAIR 1100LD Operating Manual 5 6 248083395 1 Rev 9...
Страница 146: ...Operate SOLAIR 248083395 1 Rev 9 6 82...
Страница 158: ...Lighthouse SOLAIR 1100LD Operating Manual A 2 248083395 1 Rev 9...
Страница 180: ...Lighthouse SOLAIR 1100LD Operating Manual D 2 248083395 1 Rev 9...
Страница 187: ...00...
Страница 188: ......