5-2
PCS2 Service Manual
8. Press the STOP key again to exit Configuration. Toggle the write
protect switch on the Processor Card when instructed, then replace the
rear panel.
9. Exit Utilities by pressing the STOP key.
Protocol Selection
1. Power on the PCS2 in the Utilities mode of operation by pressing and
holding the PLASMA key while switching the power on. Hold the
PLASMA key until the Utilities menu appears on the display.
2. Select Configuration on the Utilities menu by pressing the MODIFY
PROGRAM key until the arrow points to Configuration, then press
the DRAW key. Toggle the write protect switch on the Processor Card
when instructed. (Remove rear access panel.)
3. Press the MODIFY PROGRAM key until the arrow points to
Protocols.
4. Select Protocols by pressing the DRAW key.
The following graphic represents the Protocol Selection screen.
ACCESSIBLE PROTOCOLS
→∗
PPP WITHOUT SALINE COMP.
PPP WITH SALINE COMP.
PPP AND LEUKOCYTE
PPP WITH SALINE & CLOSED SET
PRESS YES/NO TO TOGGLE
Note:
Protocols with an asterisk next to them will be available to the
operator before each procedure. The PPP and Leukocyte Protocol is
used to collect leukocytes in addition to plasma. The PPP with
Saline and Closed Set Protocol is designed to be used with a
“closed set” disposable set and will return saline after each cycle.
This should not be used in the Source Plasma application.
5. Use the MODIFY PROGRAM key to move the arrow to the desired
protocol.
6. Press the YES key to select a desired protocol. (This makes the
selected protocol available to the operator.)
7. Press the NO key to deselect a desired protocol. (This makes the
deselected protocol unavailable to the operator.)
The following table has been included to aid centers in the protocol
configuration of their machines.