Chapter 6, Troubleshooting
6-31
CODE
TEXT MESSAGE
DESCRIPTION
SUGGESTED RESOLUTION
Air In Donor Line 2
Press Prime To
Continue
Fluid not seen by
DLAD2 after 40 ml
of anticoagulant
pumped.
1. Ensure proper placement of tubing in DLAD2.
2. Ensure no kinks or twists in tubing from ACAD to
DLAD2.
3. Ensure DLAD2 is clean.
4. Press PRIME to continue.
5. If fluid is present at DLAD2, use air detector test in
Diagnostics to confirm problem. Replace DLAD2 if
necessary.