Chapter 2, Machine Part Replacement Reference
2-23
when the card is mounted. If the heatsink is located toward the top
of the frame, there will not be enough slack in the harnesses to re-
connect P1, P2, and P3.
6. Mount the photoelectric assembly to the power supply plate of the
Condor power supply assembly using the same (2) Phillips head
screws.
Note:
Mount the photoelectric assembly with the harness, P405, facing
the same direction as the Condor power supply harness, P507.
7. Remove the protective label on the back of the cord clip and affix the
cord clip to the side of the cabinet above the line conditioner.
8. Partially insert the power supply assembly into the cabinet from the
rear cavity, and then align the frame with the plastic edge from the
front cavity. Align the rear brackets of the power supply with the two
threaded holes of the cabinet’s bottom panel and secure the power
supply assembly to the cabinet by using the (2) Phillips head screws.
9. Connect the 3 wire harness assembly (AC lines) on the power supply
as follows: Quickslide (blue wire) to N terminal of the line
conditioner, Quickslide (brown wire) to L terminal of the line
conditioner and attach the green wire to the GND terminal block of
the PCS2 cabinet using (1) Phillips head screw and star washer.
10. Reattach P1, P2, and P3 to the centrifuge controller card.
11. Reattach the fiber-optic cables to the photoelectric assembly by
unwrapping the cables around the centrifuge and running them
through the cord clip to the photoelectric assembly.
12 Reattach P405 to the centrifuge distribution card.
13. Reattach P507 to the backplane board.
14. Recalibrate the Condor power supply referring to the procedure in
Chapter 4, Calibration.
Replace Condor Power Supply with Condor Power Supply
Note:
Use the following directions if removing a Condor power supply
(P/N 48102-00).
1. Remove front and rear panels. (See page 2-2.)
2. Disconnect P1, P2, and P3 from the centrifuge controller card.
3. Disconnect P405 from the centrifuge distribution card and thread the
cable through the cabinet to the photoelectric assembly.
4. Disconnect P507 from the backplane card.
5. Unplug the L (brown) and N (blue) AC power lines and the Ground
wire from the back of the power supply.
6. Remove the two (2) optic cables from the photoelectric assembly by
unscrewing the two (2) plastic thumb screw fasteners.