6-16
PCS2 Service Manual
CODE
TEXT MESSAGE
DESCRIPTION
SUGGESTED RESOLUTION
206
207
208
209
210
RATIO FAULT
TOO HIGH
AC PUMP WRONG
DIRECTION
BLOOD PUMP
WRONG
DIRECTION
AC PUMP
POSITION FAULT
BLOOD PUMP
POSITION FAULT
Blood and AC pump
encoder signals
indicate that pump
ratio is higher than
set ratio.
AC pump encoder
signals indicate that
pump is moving in
wrong direction.
Blood pump encoder
signals indicate that
pump is moving in
wrong direction.
Driver board
indicates that AC
pump position is 128
counts ahead or
lagging behind set
pump position.
Driver board
indicates that blood
pump position is 128
counts ahead or
lagging behind set
pump position.
1. Swap motor and encoder plugs at top deck distribution
card to isolate problem.
2. Replace pump motor, or driver or processor card. If
CPU, recalibrate A/D, centrifuge, and safety system, and
reconfigure the language in Utilities. Reset protocol
parameters (MODIFY key), if parameters were different
from default values.
1. Swap motor and encoder plugs at top deck distribution
card to isolate problem.
2. Replace pump motor, or driver or processor card. If
CPU, recalibrate A/D, centrifuge, and safety system, and
reconfigure the language in Utilities. Reset protocol
parameters (MODIFY key), if parameters were different
from default values.
Note:
In some cases this message may be the result of flow
problems, improperly loaded harness, kinks or improperly
seated pump rotor.
1. Swap motor and encoder plugs at top deck distribution
card to isolate problem.
2. Replace pump motor, or driver or processor card. If
CPU, recalibrate A/D, centrifuge, and safety system, and
reconfigure the language in Utilities. Reset protocol
parameters (MODIFY key), if parameters were different
from default values.
1. Swap motor and encoder plugs at top deck distribution
card to isolate problem.
2. Replace pump motor, or driver or processor card. If
CPU, recalibrate A/D, centrifuge, and safety system, and
reconfigure the language in Utilities. Reset protocol
parameters (MODIFY key), if parameters were different
from default values.
1. Swap motor and encoder plugs at top deck distribution
card to isolate problem.
2. Replace pump motor, or driver or processor card. If
CPU, recalibrate A/D, centrifuge, and safety system and
reconfigure the language in Utilities. Reset protocol
parameters (MODIFY key), if parameters were different
from default values.