Cleaning Supplies Needed
1. Cleaning solution (recommended for use with bloodborne pathogens)
2. Clear, warm water
3. Distilled water
4. Lint-free gauze pads or towels
5. Protective gloves
6. Phillips head screwdriver
7. Silicone vacuum grease
8. Cotton swabs
9. Spray bottles
10. Owner’s Manual and/or Service Manual
11. Maintenance records or documentation logs
Warning!
Power off and unplug the PCS2 before cleaning to prevent
potential danger of electrical shock.
Note:
Throughout this chapter, some cleaning procedures reference using
distilled water to clean components. We suggest using distilled
water as it is less likely to leave a film on the components.
However, if distilled water is not available, use the clearest water
available at your center.
Cleaning the PCS2
Note: Certain commercial cleaning agents may not be compatible with
components on the PCS2. Several commonly used agents known
NOT to be compatible with the PCS2 are Pro-Tech, Dispatch,
CitriGuard II and Alcohol. If there are questions regarding the use
of a specific cleaning agent, call the Haemonetics Hotline with the
Manufacturer's Information on the cleaning agent and the cleaning
agents Material Safety Data Sheet (MSDS).
Membrane Panel and Outer Cabinet
The exterior of the PCS2, including the membrane panel, should be
cleaned with a cleaning solution at regular intervals as well as any time a
spill has occurred. Do not spray cleaning solution directly on the
membrane panel.
8-2
PCS2 Service Manual
!