6-10
PCS2 Service Manual
CODE
TEXT MESSAGE
DESCRIPTION
SUGGESTED RESOLUTION
32
33
34
36
37
38
39
CHANNEL 1
STATE ERROR
CHANNEL 1
STATE ERROR
CHANNEL 2
STATE ERROR
(AIR STATE TEST)
CHANNEL 2
STATE ERROR
CHANNEL 2
STATE ERROR
CHANNEL 2
STATE ERROR
CHANNEL 2
STATE ERROR
(AIR STATE TEST)
Safety system relay
did not close during
system start-up
Channel 1 state test.
Safety system did not
post fault to register
during system start-
up Channel 1 state
test.
Safety system did not
respond to air signal
during system start-up
Channel 1 and 2 air
detector state test.
Safety system relay
did not open during
system start-up
Channel 2 state test.
Safety system relay
did not close during
system start-up
Channel 2 state test.
Safety system did not
post fault to register
during system start-
up Channel 2 state
test.
Safety system relay
did not open during
system start-up
Channel 1 and 2 air
detector state test.
1. Replace driver card.
2. Replace safety card.
1. Replace safety card.
2. Replace processor card. Using Utilities, calibrate all
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Replace safety card.
1. Replace driver card.
2. Replace safety card.
1. Replace driver card.
2. Replace safety card.
1. Replace safety card.
2. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Replace driver card.
2. Replace safety card.