Chapter 6, Troubleshooting
6-11
CODE
TEXT MESSAGE
DESCRIPTION
SUGGESTED RESOLUTION
40
41
43
44
45
47
48
CHANNEL 2
STATE ERROR
(AIR STATE TEST)
CHANNEL 2
STATE ERROR
(AIR STATE TEST)
CHANNEL 1 TIME
BASE ERROR
CHANNEL 1 TIME
BASE ERROR
CHANNEL 1 TIME
BASE ERROR
CHANNEL 2 TIME
BASE ERROR
CHANNEL 2 TIME
BASE ERROR
Safety system relay
did not close during
system start-up
Channel 1 and 2 air
detector state test.
Safety system did not
post fault to register
during system start-up
Channel 1 and 2 air
detector state test.
Safety system relay
did not open during
system start-up
Channel 1 time base
test.
Safety system relay
did not close during
system start-up
Channel 1 time base
test.
Safety system did not
post fault to register
during system start-
up Channel 1 time
base test.
Safety system relay
did not open during
system start-up
Channel 2 time base
test.
Safety system relay
did not close during
system start up
Channel 2 time base
test.
1. Replace driver card.
2. Replace safety card.
1. Replace safety card.
2. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Replace driver card.
2. Replace safety card.
1. Replace driver card.
2. Replace safety card.
1. Replace safety card.
1. Replace driver card.
2. Replace safety card.
1. Replace driver card.
2. Replace safety card.